Agilent 8800 Triple Quadrupole ICP-MS – Site Preparation Checklist

Manuals | 2015 | Agilent TechnologiesInstrumentation
ICP/MS, ICP/MS/MS
Industries
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


Proper site preparation for a high performance ICP triple quadrupole mass spectrometer is essential to ensure long term reliability, optimal analytical performance and operator safety. Defining the required laboratory space, environmental controls, power supply, cooling systems and gas delivery infrastructure minimizes installation delays, protects instrument integrity and enhances data quality in trace elemental analysis.

Objectives and Overview of the Document


This site preparation checklist aims to provide a consolidated guide for laboratory managers and technicians to verify that all facility, utility and equipment requirements for the Agilent 8800 Triple Quadrupole ICP-MS are met prior to delivery and installation. The document covers:
  • Laboratory bench dimensions and load capacity
  • Ambient conditions, exhaust and ventilation requirements
  • Electrical power specifications and grounding
  • Cooling water supply and quality criteria
  • High purity gas connections and regulators
  • Network connectivity and other ancillary needs

Methodology and Used Instrumentation


The checklist is organized by major utility and spatial categories, referencing both standard laboratory infrastructure and Agilent supplied modules. The primary instrumentation includes the Agilent 8800 Triple Quadrupole ICP-MS mainframe, foreline pump, PC workstation, chiller or heat exchanger, and optional autosamplers. Supporting components such as gas regulators, stainless steel tubing and base plates are specified to maintain instrument performance.

Main Requirements and Discussion


Key site preparation parameters are summarized below:
  • Space and Access: A clear work envelope of at least 60 cm around the instrument with bench support for up to 382 kg shipping weight and 316 kg mainframe.
  • Environmental Conditions: Stable ambient temperature between 15–30 °C with less than 2 °C/hour drift, non-condensing humidity (20–80 %), maximum altitude 2000 m, and a dedicated exhaust flow of 5–7 m3/min through 150 mm ducting.
  • Electrical Supply: Isolated ground, 200–240 Vac at 50/60 Hz with 24 A rating for the mass spectrometer, plus outlets for PC, chiller, heat exchanger and autosamplers.
  • Cooling Water: Flow rate above 5 L/min at inlet temperature of 15–40 °C, conductivity 50–150 µS, filtered to 100 µm particle size, preferably via an Agilent chiller or approved recirculator.
  • Gas Delivery: High purity argon (≥99.99 %) at 500–700 kPa for plasma, and cell gases (He, H2, NH3/He, O2) at ≥99.999 % purity, regulated and plumbed using Agilent stainless steel tubing under 200 kPa, within 3 m of the instrument.
  • Network and IT: Optional LAN connection via a dedicated switch, two static IP addresses and no dual network cards on the instrument PC.

Benefits and Practical Applications


Adhering to these preparation guidelines ensures a smooth installation process, reduces the risk of instrument downtime, and maximizes analytical throughput. Proper utility sizing and layout protects sensitive optics and electronics, while dedicated exhaust and gas controls safeguard personnel and maintain compliance with laboratory safety standards.

Future Trends and Opportunities


Advances in laboratory automation and digital integration are driving more compact and modular ICP-MS installations. Emerging remote monitoring systems and predictive maintenance analytics will further streamline site qualification, enabling rapid deployment in quality control, environmental monitoring and research laboratories. Sustainable utilities and green solvent handling represent additional growth areas for future installations.

Conclusion


Comprehensive site preparation is the foundational step for achieving reliable operation and high quality results from the Agilent 8800 Triple Quadrupole ICP-MS. By verifying space, environmental, power, cooling and gas supply requirements in advance, laboratories can minimize installation risks and accelerate their analytical workflows.

Reference


  • Agilent Technologies. Agilent 8800 Triple Quadrupole ICP-MS Site Preparation Checklist. Document G3660-90064, Revision 1.3, July 2015.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
7700 Series ICP-MS – Site Preparation Checklist
7700 Series ICP-MS – Site Preparation Checklist
2015|Agilent Technologies|Manuals
7700 Series ICP-MS – Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the…
Key words
checklist, checklistsite, sitespecial, specialnotes, notestechnologies, technologiescompressed, compressedagilent, agilentgas, gaspreparation, preparationwatts, wattsconsumption, consumptionrequirements, requirementsheat, heatdescription, descriptionkpa
7700 Series ICP-MS – Site Preparation Checklist
7700 Series ICP-MS – Site Preparation Checklist
2013|Agilent Technologies|Manuals
7700 Series ICP-MS – Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the…
Key words
checklist, checklistsite, sitespecial, specialnotes, notestechnologies, technologiescompressed, compressedagilent, agilentgas, gaspreparation, preparationwatts, wattsconsumption, consumptionrequirements, requirementsdescription, descriptionkpa, kpaoperating
1 700 Series ICP-OES Site Preparation Checklist
1 700 Series ICP-OES Site Preparation Checklist
2012|Agilent Technologies|Manuals
700 Series ICP-OES Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the key…
Key words
checklist, checklistsite, sitespecial, specialtechnologies, technologiesnotes, noteswatts, wattsagilent, agilentpreparation, preparationoperating, operatingcustomer, customerconsumption, consumptioncompressed, compressedheat, heatinstrument, instrumentweight
Agilent 5100 Series ICP-OES Site Preparation Checklist
Agilent 5100 Series ICP-OES Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the…
Key words
checklist, checklistsite, sitespecial, specialtechnologies, technologiesnotes, notesagilent, agilentwatts, wattspreparation, preparationoperating, operatingcustomer, customerdescription, descriptionmaximum, maximumconsumption, consumptioncompressed, compressedheat
Other projects
GCMS
LCMS
Follow us
FacebookLinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike