7700 Series ICP-MS – Site Preparation Checklist

Manuals | 2013 | Agilent TechnologiesInstrumentation
ICP/MS
Industries
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


The Agilent 7700 Series ICP-MS requires careful site preparation to ensure reliable performance and long instrument life. Proper laboratory space, utilities, and environmental controls are critical to achieve consistent analytical results and avoid installation delays.

Objectives and Study Overview


This document provides a comprehensive pre-installation checklist outlining the necessary laboratory bench dimensions, environmental conditions, power and cooling requirements, gas supplies, and safety considerations to facilitate a smooth and timely setup of the 7700 Series ICP-MS system.

Methodology and Instrumentation


  • Laboratory Space and Access: Reserve bench space capable of supporting containers up to 157 kg (347 lb) and the mainframe at 115 kg (254 lb), with 60 cm (2 ft) clearance on all sides for maintenance. Ensure delivery routes accommodate the largest shipping container and provide a forklift or lifting device.
  • Environmental Conditions: Maintain ambient temperature between 15–30 °C (59–86 °F) with changes under 2 °C per hour and total variation below 5 °C. Control humidity at 20–80% non-condensing. Install continuous exhaust flow of 5–7 m³/min through 150 mm ducting to vent plasma and pump emissions externally; fluctuation must remain within ±5%.
  • Power Requirements: Provide dedicated outlets matching instrument specifications (e.g., 200–240 VAC, 50/60 Hz, 24 A for ICP-MS; additional outlets for PC, chiller, autosampler). Use an isolated, noise-free ground and avoid extension cords; install local regulators or twist-lock plugs as required by region.
  • Cooling Water System: Prefer Agilent chiller (G3292A) or heat exchanger (G1879B) charged with Poly-Clear fluid. If using alternative recirculators, fill with distilled water (50–150 μS conductivity) at 15–40 °C inlet temperature and supply >5 L/min at 230–400 kPa inlet pressure. Include ≤100 μm inline filters.
  • Gas Supplies: Install high-purity Argon (≥99.99%) at 500–700 kPa for plasma gas and cell gas options (He, H₂ or NH₃/He mix at ≥99.999%) with dedicated two-stage regulators, stainless steel tubing under 3 m, and proper fittings. Avoid shared or house gas lines to prevent contamination.

Key Results and Discussion


Strict adherence to these specifications directly impacts instrument performance, stability of plasma, and data quality. Adequate space and clearances prevent mechanical interference, stable environmental conditions protect sensitive components, and proper utilities ensure uninterrupted operation. Failure to meet any requirement can lead to rescheduled installations, increased downtime, or compromised analytical results.

Benefits and Practical Applications


Implementing a detailed site preparation strategy reduces commissioning time, minimizes maintenance interventions, and safeguards investment. Laboratories engaged in environmental testing, pharmaceutical analysis, materials characterization, and QA/QC benefit from consistent instrument performance and reproducible data.

Future Trends and Potential Applications


Advancements in pre-installation software assessment, remote readiness verification, and sustainable utility solutions (e.g., energy-efficient chillers and low-impact exhaust systems) will streamline ICP-MS deployment. Integration with smart building controls and IoT monitoring promises further optimization of lab infrastructure for analytical workflows.

Conclusion


Comprehensive site preparation is essential for leveraging the high sensitivity, robustness, and throughput of the Agilent 7700 Series ICP-MS. By following this checklist, laboratories ensure a smooth installation, reliable operation, and long-term analytical excellence.

References


  • Agilent Technologies. 7700 Series ICP-MS Site Preparation Checklist. Document G3280-90064, May 2013.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
7700 Series ICP-MS – Site Preparation Checklist
7700 Series ICP-MS – Site Preparation Checklist
2015|Agilent Technologies|Manuals
7700 Series ICP-MS – Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the…
Key words
checklist, checklistsite, sitespecial, specialnotes, notestechnologies, technologiescompressed, compressedagilent, agilentgas, gaspreparation, preparationwatts, wattsconsumption, consumptionrequirements, requirementsheat, heatdescription, descriptionkpa
Agilent 8800 Triple Quadrupole ICP-MS – Site Preparation Checklist
Agilent 8800 Triple Quadrupole ICP-MS – Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation…
Key words
checklist, checklistsite, sitespecial, specialnotes, notestechnologies, technologiescompressed, compressedagilent, agilentgas, gaspreparation, preparationwatts, wattsconsumption, consumptionrequirements, requirementsheat, heatdescription, descriptionkpa
1 700 Series ICP-OES Site Preparation Checklist
1 700 Series ICP-OES Site Preparation Checklist
2012|Agilent Technologies|Manuals
700 Series ICP-OES Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the key…
Key words
checklist, checklistsite, sitespecial, specialnotes, notestechnologies, technologieswatts, wattsagilent, agilentpreparation, preparationoperating, operatingcustomer, customerconsumption, consumptioncompressed, compressedheat, heatinstrument, instrumentweight
Agilent 5100 Series ICP-OES Site Preparation Checklist
Agilent 5100 Series ICP-OES Site Preparation Checklist Thank you for purchasing an Agilent instrument. To get you started and to assure a successful and timely installation, please refer to this specification or set of requirements. Correct site preparation is the…
Key words
checklist, checklistsite, sitespecial, specialnotes, notestechnologies, technologiesagilent, agilentwatts, wattspreparation, preparationoperating, operatingcustomer, customerdescription, descriptionmaximum, maximumconsumption, consumptioncompressed, compressedheat
Other projects
GCMS
LCMS
Follow us
FacebookLinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike