K-Alpha: Energy Scale Linearity and Calibration
Applications | 2008 | Thermo Fisher ScientificInstrumentation
The accurate calibration and linearity of the energy scale in X-ray photoelectron spectroscopy (XPS) are critical for reliable identification of chemical states. In industrial and research settings, consistent binding energy measurements underpin material characterization and quality control.
This study evaluates the calibration routine and energy scale stability of the Thermo Scientific K-Alpha XPS system. It assesses linearity across multiple reference peaks and examines short-term stability through repeated measurements over two days.
Calibration employs built-in gold (Au 4f7/2), silver (Ag 3d5/2), and copper (Cu 2p3/2) standards mounted in the sample stage. Each peak is scanned over a defined energy range with 50 meV step size. Peak maxima are determined by parabolic least-squares fitting of the upper intensity region. The automated Avantage software routine adjusts gain and offset to align measured energies with reference values. For stability assessment, each standard is measured 100 times under identical conditions without recalibration.
The automated routine delivered mean binding energy errors of 6 meV and a standard deviation of 16 meV (3σ ≈ 48 meV), closely matching the measurement step size. The total spread in energy was marginally above 50 meV. Calculated effective photon energies averaged 1486.69 eV with a standard deviation of 24 meV, all within ISO 15472 limits (1486.61–1486.81 eV). No systematic drift or trend was observed, indicating robust monochromator alignment and energy scale linearity.
Future developments may include integration of remote monitoring and cloud-based calibration logs, AI-driven diagnostics for predictive maintenance, enhanced monochromator designs for improved photon energy stability, and wider adoption in regulated environments requiring stringent traceability.
The K-Alpha automated calibration routine offers a fast, efficient, and reliable method to maintain energy scale accuracy and stability. Its built-in standards and software controls enable high-confidence data acquisition without interrupting user workflows, meeting rigorous QA/QC demands in research and industrial laboratories.
X-ray
IndustriesMaterials Testing
ManufacturerThermo Fisher Scientific
Summary
Significance of the topic
The accurate calibration and linearity of the energy scale in X-ray photoelectron spectroscopy (XPS) are critical for reliable identification of chemical states. In industrial and research settings, consistent binding energy measurements underpin material characterization and quality control.
Aims and overview of the study
This study evaluates the calibration routine and energy scale stability of the Thermo Scientific K-Alpha XPS system. It assesses linearity across multiple reference peaks and examines short-term stability through repeated measurements over two days.
Methods
Calibration employs built-in gold (Au 4f7/2), silver (Ag 3d5/2), and copper (Cu 2p3/2) standards mounted in the sample stage. Each peak is scanned over a defined energy range with 50 meV step size. Peak maxima are determined by parabolic least-squares fitting of the upper intensity region. The automated Avantage software routine adjusts gain and offset to align measured energies with reference values. For stability assessment, each standard is measured 100 times under identical conditions without recalibration.
Instrumentation used
- Thermo Scientific K-Alpha XPS spectrometer
- Avantage data system for automated control and logging
- Built-in Au, Ag, Cu calibration standards
- Argon ion sputtering for sample cleaning
- Monochromator for X-ray photon energy selection
Main results and discussion
The automated routine delivered mean binding energy errors of 6 meV and a standard deviation of 16 meV (3σ ≈ 48 meV), closely matching the measurement step size. The total spread in energy was marginally above 50 meV. Calculated effective photon energies averaged 1486.69 eV with a standard deviation of 24 meV, all within ISO 15472 limits (1486.61–1486.81 eV). No systematic drift or trend was observed, indicating robust monochromator alignment and energy scale linearity.
Benefits and practical applications
- Rapid (few minutes), unattended calibration fits seamlessly into routine workflows
- Automated logging ensures traceability for QA/QC requirements
- High accuracy and precision support reliable chemical state analysis
- Built-in standards eliminate sample swapping and reduce downtime
Future trends and applications
Future developments may include integration of remote monitoring and cloud-based calibration logs, AI-driven diagnostics for predictive maintenance, enhanced monochromator designs for improved photon energy stability, and wider adoption in regulated environments requiring stringent traceability.
Conclusions
The K-Alpha automated calibration routine offers a fast, efficient, and reliable method to maintain energy scale accuracy and stability. Its built-in standards and software controls enable high-confidence data acquisition without interrupting user workflows, meeting rigorous QA/QC demands in research and industrial laboratories.
References
- International Standard ISO 15472, Surface chemical analysis – X-ray photoelectron spectrometers – Calibration of energy scales (2001)
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
K-Alpha: Mapping of the Work Function of a Damaged Solar Cell
2011|Thermo Fisher Scientific|Applications
Application Note: 52078 K-Alpha: Mapping of the Work Function of a Damaged Solar Cell Paul Mack, Tim Nunney, John Wolstenholme, Thermo Fisher Scientific, East Grinstead, UK Key Words • K-Alpha • Mapping • PCA The Thermo Scientific K-Alpha XPS has…
Key words
alpha, alphawork, workfunction, functionxps, xpssolar, solardamaged, damagedpca, pcaphoton, photonindium, indiumenergy, energymaps, mapsmeasured, measuredposition, positionafrica, africascientific
Instrumentation for surface analysis
2021|Thermo Fisher Scientific|Brochures and specifications
Instrumentation for surface analysis Surface chemistry and thin film characterization X-ray photoelectron spectroscopy Quantitative, chemical identification of the surface X-ray Photoelectron Spectroscopy (XPS, also known as Electron Spectroscopy for Chemical Analysis – ESCA) is a highly surface-sensitive, quantitative, chemical analysis…
Key words
xps, xpsspectroscopy, spectroscopymagcis, magcisescalab, escalabqxi, qximonatomic, monatomicsource, sourcesurface, surfaceion, iondepth, depthreels, reelsmicroprobe, microprobeenergy, energyfinancing, financingavantage
K-Alpha: A New Concept in XPS
2008|Thermo Fisher Scientific|Applications
Application Note: 31094 Key Words • Surface Analysis • Automation • Integrated Design • Monochromated, Small Spot XPS • Optimized Geometry K-Alpha: A New Concept in XPS Thermo Scientific K-Alpha has been designed to maximize the throughput and efficiency of…
Key words
view, viewreflex, reflexposition, positionplatter, platteralpha, alphaanalysis, analysisheight, heightgun, guncamera, cameraloadlock, loadlockillumination, illuminationflood, floodlive, liveavantage, avantagesample
Nexsa G2 Surface Analysis System
2022|Thermo Fisher Scientific|Brochures and specifications
Datasheet Nexsa G2 Surface Analysis System Fast travel to definitive surface analysis As a materials researcher, you need a surface analysis system that quickly and efficiently provides high-quality, accurate data collection so you can understand the composition of surfaces, thin…
Key words
xps, xpsalignment, alignmentholder, holdersource, sourcesystem, systemsample, samplemounting, mountingbacking, backingsoftware, softwareixr, ixrholders, holdersmaps, mapsdata, dataavantage, avantagespot