High precision hafnium isotope ratio measurements

Applications | 2021 | Thermo Fisher ScientificInstrumentation
ICP/MS
Industries
Environmental
Manufacturer
Thermo Fisher Scientific

Summary

Significance of the Topic


Precise hafnium isotope analysis plays a critical role in geosciences by providing insights into planetary differentiation, crustal evolution and mantle processes. Enhanced sensitivity and precision in these measurements directly improve the reliability of isotopic data for age dating and provenance studies.

Study Objectives and Overview


This application brief evaluates the performance of the Thermo Scientific Neoma MC-ICP-MS for high precision hafnium isotope ratio determination. The study focuses on assessing sensitivity gains, isotope ratio precision and accuracy, as well as throughput potential for routine geochemical applications.

Methodology and Instrumentation


  • Sample Preparation A 200 ppb hafnium solution was introduced using a 100 µL/min self aspirating nebulizer and an SIS spray chamber.
  • Measurement Protocol Ten sequential blocks of 10 minute integration were performed, with internal normalization of 179Hf/177Hf by the exponential mass bias correction model.
  • Interference Corrections Isobaric overlaps from ytterbium, lutetium, tantalum and tungsten were corrected mathematically.

Used Instrumentation


  • Instrument Thermo Scientific Neoma MC-ICP-MS high performance multi collector ICP-MS.
  • Detector Array Eleven Faraday cups equipped with automated alignment and 1011 Ohm amplifiers.
  • Mass Resolving Power Approximately 1950 across all detector cups.

Main Results and Discussion


  • Sensitivity The system achieved 5.23 gigacounts per second per ppm, equivalent to 83.7 volts per ppm, nearly double that of previous generation instruments under wet plasma conditions.
  • Precision and Accuracy The mean 176Hf/177Hf ratio fell within the accepted standard window of –25 to +55 ppm. Reported RSD values for 176Hf/177Hf, 178Hf/177Hf and 180Hf/177Hf were all below 10 ppm over ten replicates.
  • Stability Individual standard errors reached the theoretical limit set by counting statistics, demonstrating excellent measurement stability over extended runs.

Benefits and Practical Applications


  • Routine Workflow The enhanced sensitivity allows faster sample throughput and reduced analysis times.
  • Sample Conservation Lower sample consumption broadens the study of trace-level hafnium in limited materials.
  • High Precision Geochemical Chronologies Improved isotope precision supports more accurate age models and provenance tracing.

Future Trends and Opportunities


Ongoing developments may target higher detector resolution, automated sample handling and expanded application to other isotopic systems. Potential fields of use include environmental monitoring, nuclear forensics and advanced materials characterization.

Conclusion


The Neoma MC-ICP-MS platform delivers a significant advancement in hafnium isotope ratio measurement by doubling sensitivity and achieving sub 10 ppm precision. Its performance and throughput capabilities make it a powerful tool for high accuracy geochemical research.

Reference


  • No additional references were provided in the original document.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
High precision 176Hf/177Hf measurements in zircons by LA-MC-ICP-MS
APPLICATION NOTE High precision 176Hf/177Hf measurements in zircons by LA-MC-ICP-MS Authors: Grant Craig, Jenny Roberts, Claudia Bouman, Nicholas S. Lloyd, and Johannes B. Schwieters; Thermo Fisher Scientific, Bremen, Germany Keywords: Neoma, MC-ICP-MS, Hafnium, isotope ratio, Laser Ablation, Geosciences Introduction The…
Key words
icp, icpneoma, neomaplešovice, plešoviceablation, ablationqtegra, qtegrazircons, zirconslaser, lasermud, mudisotope, isotopeisds, isdstank, tanknwr, nwrjet, jetplugins, pluginsplugin
Performance of Neoma MS/MS MC-ICP-MS for analysis of radiogenic isotopes
Application note | 002361 Mass spectrometry Performance of Neoma MS/MS MC-ICP-MS for analysis of radiogenic isotopes Authors Goal A. Santato, G. Craig, To show performance of Thermo Scientific™ Neoma™ MS/MS MC-ICP-MS for analysis of C. Bouman, J. Roberts non-traditional radiogenic…
Key words
neoma, neomaslot, slotbias, biasmerck, merckexponential, exponentialprecision, precisionicp, icpradiogenic, radiogenichafnium, hafniumneodymium, neodymiummass, masscorrections, correctionsstrontium, strontiumisotope, isotopecounting
Neoma Multicollector ICP-MS
Neoma Multicollector ICP-MS
2020|Thermo Fisher Scientific|Brochures and specifications
Illuminating the edge of discovery Neoma Multicollector ICP-MS Illuminating the edge of discovery Just when you thought it couldn’t be done, the Thermo Scientific™ Neoma™ MC-ICP-MS raises the bar. Building on the strength of our multicollector technology, this next generation…
Key words
srf, srfneoma, neomaicp, icpfaraday, faradaymass, massplasma, plasmacps, cpsmulticollector, multicollectorrpq, rpqcups, cupsamplifier, amplifierwet, wetgcps, gcpscounting, countingsem
High precision δ41K using MC-ICP-MS in high resolution and MS/MS mode
Application note | 001035 Mass spectrometry High precision δ41K using MC-ICP-MS in high resolution and MS/MS mode Authors Goal G. Craig, M. Pfeifer, C. Bouman, To demonstrate how the different technologies within the Thermo Scientific™ Neoma™ J. Roberts, N. Lloyd,…
Key words
neoma, neomaxhr, xhricp, icpoption, optiontorch, torchcps, cpsposition, positioncrc, crcauxilary, auxilaryomega, omegamicrofast, microfastapex, apexcater, catergas, gasmin
Other projects
GCMS
LCMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike