ICPMS
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike

Non-Destructive Analysis of Substrates and Contaminants by FTIR with Specular Reflectance Interface

Applications | 2023 | Agilent TechnologiesInstrumentation
FTIR Spectroscopy
Industries
Materials Testing
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


The cleanliness of a substrate surface is critical in industrial and engineering applications where bonding or coating is required. Residual oils or release agents can reduce bonding strength, impair adhesion, and lead to premature failure of assemblies. A rapid, non-destructive in-situ technique for detection and quantification of surface contaminants supports reliable quality control and process optimization.

Objectives and Study Overview


This application note demonstrates the capabilities of the Agilent 4300 handheld FTIR spectrometer equipped with a 45° specular reflectance interface. The goals were to verify the instrument’s ability to:
  • Identify substrate materials
  • Detect the presence and type of surface contaminants
  • Quantify contaminant levels
  • Confirm chemical condition of surface agents

Experiments included qualitative analyses on a glass-lined vessel smeared with various lubricants and quantitative studies of PTFE-based release agent deposits on aluminum coupons.

Methodology and Instrumentation


The Agilent 4300 FTIR handheld spectrometer was fitted with a specular reflectance sampling interface optimized for reflective and semi-reflective surfaces. Key methodological points:
  • Interface selection considered ATR, transmission, diffuse reflectance and specular reflectance. Specular reflectance at a 45° cone half-angle was chosen for hard, opaque substrates.
  • Qualitative samples: clean glass vessel wiped, then smeared with six types of lubricants or release agents; spectra acquired at 8 cm–1 resolution with 64 scans (~16 s per measurement).
  • Quantitative samples: aluminum coupons (40.85 × 50.80 mm) contaminated with precisely weighed PTFE aerosol release agent (0–10.28 mg); clean coupon prepared by ethanol wash and lint-free drying.
  • Data processing: spectral preprocessing minimal; multivariate calibration using partial least squares (PLS1) in Agilent MicroLab Expert software with a 4:1 calibration/validation split. Both global and low-range models were developed.
  • Method deployment: validated models transferred to MicroLab Mobile for on-site operation with step-by-step guidance and traffic-light result display.

Main Results and Discussion


Qualitative spectra of six lubricants on a glass-lined vessel exhibited distinct CH-stretch and fingerprint regions, enabling straightforward identification. Infrared reflection-absorption (IRRAS) doubled path length through thin films, enhancing sensitivity. Quantitative PLS1 models achieved:
  • High linearity (R2 > 0.98) and slope ~0.98 over 0–0.495 mg/cm2 PTFE
  • Limit of detection down to ~0.004 mg/cm2
  • Reproducible predictions confirmed by boxplot analysis of replicate spectra

Minimal preprocessing was required compared to earlier studies on polymer composites. The method can distinguish physisorbed water and native oxide signals from true contaminants.

Benefits and Practical Applications


  • Non-destructive, force-free “point and shoot” measurements
  • Rapid analysis (< 30 s per result) with clear on-screen interpretation
  • Portable handheld design (~2 kg) suitable for field and production-line use
  • Applicable to metals, alloys, glass, ceramics, composites, textiles, concrete and more
  • Facilitates real-time quality assurance of bonding, coating and mold-release processes

Future Trends and Opportunities


Advances may include integration with automated spectral libraries and AI-driven identification, real-time process control feedback, expanded multivariate models for multiple contaminants, and remote monitoring via wireless connectivity. Further applications could target trace-level detection in pharmaceutical or semiconductor manufacturing.

Conclusion


The Agilent 4300 handheld FTIR spectrometer with specular reflectance interface offers a versatile, sensitive, and rapid solution for in-situ surface cleanliness testing. Both qualitative contaminant identification and quantitative measurement of PTFE release agents on aluminum were achieved with high accuracy and minimal sample preparation. The system supports robust quality control in critical bonding and coating workflows.

Reference


  1. Extended Non-Destructive Testing of Composite Bonds (ENCOMB), FP7 CORDIS, European Commission (europa.eu), accessed April 2023.
  2. Tang, P. L.; Rein, A. Quantification of Release Agent on a Carbon-Fiber-Reinforced Polymer using a Handheld FTIR, Agilent Technologies Application Note 5991-5595EN, 2018.
  3. Tang, P. L. Comparison of Portable FTIR Interface Technologies for the Analysis of Paints, Minerals & Concrete, Agilent Technologies Application Note 5991-8359EN, 2017.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Molecular Spectroscopy Application eHandbook
Molecular Spectroscopy Application eHandbook
2017|Agilent Technologies|Guides
Home Previous Next TABLE OF CONTENTS COATING CHALLENGES INSTRUMENT OVERVIEW + MEASUREMENTS OF COATINGS OTHER COATING TECHNOLOGIES GATHER RICH INSIGHTS FROM COATINGS ANALYSIS Molecular Spectroscopy Application eHandbook Home Previous Next TABLE OF CONTENTS COATING CHALLENGES INSTRUMENT OVERVIEW + MEASUREMENTS OF…
Key words
ftir, ftircoating, coatingcoatings, coatingsmeasurements, measurementsreinforced, reinforcedpet, petanodization, anodizationthickness, thicknessaluminum, aluminumfiber, fibereasuring, easuringhome, homereflectance, reflectanceidentification, identificationndt
Fourier Transform Infrared Spectroscopy for Rapid Cleaning Verification of Mixing Vessels and Reaction Chambers
Application Note Biopharma and Pharma Fourier Transform Infrared Spectroscopy for Rapid Cleaning Verification of Mixing Vessels and Reaction Chambers Authors David Troiani, PhD., Marco Pinto, PhD., and Alex Harvey, PhD. Agilent Technologies, Inc. Abstract Rapid cleaning verification using Fourier transform…
Key words
coupons, couponssurface, surfacecleaning, cleaningmodels, modelscoupon, couponftir, ftirvessel, vesselreflectance, reflectancecorrelating, correlatingsteel, steelconcentration, concentrationdot, dotcontaminant, contaminantakin, akinpredicted
Quantification of Release Agent on a Carbon-Fiber-Reinforced Polymer using a Hand-Held FTIR
Application Note Materials Research & Testing Quantification of Release Agent on a Carbon-Fiber-Reinforced Polymer using a Hand-Held FTIR Non-destructive, in-situ analysis completed in less than one minute Authors Pik Leung Tang, Alan Rein Agilent Technologies Introduction Release agents are applied…
Key words
ftir, ftircfrp, cfrptreatment, treatmentbond, bondgic, gicagent, agentseverity, severitystrength, strengthrelease, releasexps, xpsscenarios, scenariosmodel, modelhandheld, handheldlevels, levelsspectroscopy
Analysis of plasma treated carbon fiber reinforced polymer (CFRP) composites by portable Fourier Transform Infrared Spectroscopy (FTIR)
Analysis of plasma treated carbon fiber reinforced polymer (CFRP) composites by portable Fourier Transform Infrared Spectroscopy (FTIR) Application note Materials testing Authors Alan Rein, Ph.D. Pik Leung Tang, Ph.D. Agilent Technologies, Inc., USA Introduction Verified for Agilent 4300 Handheld FTIR…
Key words
cfrp, cfrpnozzle, nozzletreated, treatedtreatment, treatmentplasma, plasmacoupon, couponpdms, pdmsftir, ftirpeel, peelsurface, surfacewax, waxheight, heighthandheld, handheldcomposite, compositerelease
Other projects
GCMS
LCMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike