The Agilent 7700x/7800 ICP-MS Advantage for Drinking Water Analysis
Brochures and specifications | 2015 | Agilent TechnologiesInstrumentation
The accurate determination of trace metals in drinking water is critical for public health, regulatory compliance and environmental monitoring. Polyatomic interferences commonly encountered in ICP-MS analyses can compromise detection limits and data quality. The introduction of an optimized collision/reaction cell (ORS) capable of universal interference removal simplifies method development and improves performance across a wide range of sample chemistries.
This application note evaluates the performance of the Agilent 7700x/7800 Series ICP-MS equipped with the Octopole Reaction System (ORS) in helium collision mode for routine analysis of trace metals in drinking waters. Key goals include:
The 7700x/7800 ICP-MS incorporates:
Calibration standards contained 1% HNO3 and 0.5% HCl to stabilize elements such as Ag, Sb and Hg. Helium mode effectively removed chloride-based interferences on As, Se, Cr and V without resorting to reactive gases.
Representative calibration demonstrated linearity (R² ~1.000) and low background equivalent concentrations, for example a 3.7 ppt BEC for vanadium after ClO+ removal. Method detection limits (MDLs) determined as three sigma of ten replicates under default plasma conditions include:
A continuous 11-hour sequence of 190 samples (municipal waters, two high-TDS mineral waters, and a diluted NIST 1643e reference) yielded:
The combination of ORS, advanced ion optics and detector delivers:
These advantages support routine drinking water monitoring, regulatory compliance testing and broader environmental or industrial analyses without instrument reconfiguration.
Further developments may include:
The Agilent 7700x/7800 ICP-MS with ORS in helium mode provides a robust, high-sensitivity solution for trace metal analysis in drinking waters. Universal interference removal, low detection limits and simplified operation yield reliable data quality and high productivity. The system’s wide dynamic range and HMI capability ensure consistent performance across a variety of sample matrices without the need for reactive cell gases or complex corrections.
Steve Wilbur. Agilent 7700x/7800 Series ICP-MS Octopole Reaction System for Drinking Water Analysis. Agilent Technologies Application Note 5990-4315EN, July 2015.
ICP/MS
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Importance of the topic
The accurate determination of trace metals in drinking water is critical for public health, regulatory compliance and environmental monitoring. Polyatomic interferences commonly encountered in ICP-MS analyses can compromise detection limits and data quality. The introduction of an optimized collision/reaction cell (ORS) capable of universal interference removal simplifies method development and improves performance across a wide range of sample chemistries.
Study objectives and overview
This application note evaluates the performance of the Agilent 7700x/7800 Series ICP-MS equipped with the Octopole Reaction System (ORS) in helium collision mode for routine analysis of trace metals in drinking waters. Key goals include:
- Assessing the ease of operation under a single set of conditions without prior sample knowledge
- Demonstrating improved detection limits and interference removal, particularly for iron and selenium
- Verifying robustness over extended sequences of high-matrix samples
Methodology and instrumentation
The 7700x/7800 ICP-MS incorporates:
- An optimized ORS for helium collision mode, enabling universal removal of polyatomic interferences
- Improved ion lens design and a new detector, providing higher sensitivity and lower background
- One-click Plasma Setting and Expert AutoTune for fully automated, reproducible instrument optimization
- High Matrix Introduction (HMI) system for stable, long-term analysis of samples up to several hundred mg/L TDS
Calibration standards contained 1% HNO3 and 0.5% HCl to stabilize elements such as Ag, Sb and Hg. Helium mode effectively removed chloride-based interferences on As, Se, Cr and V without resorting to reactive gases.
Main results and discussion
Representative calibration demonstrated linearity (R² ~1.000) and low background equivalent concentrations, for example a 3.7 ppt BEC for vanadium after ClO+ removal. Method detection limits (MDLs) determined as three sigma of ten replicates under default plasma conditions include:
- Fe and Se MDLs below 20 ppt in helium mode
- Be and B MDLs around 5 ppt in no-gas mode
- Most other elements with MDLs between 2 and 15 ppt in helium mode
A continuous 11-hour sequence of 190 samples (municipal waters, two high-TDS mineral waters, and a diluted NIST 1643e reference) yielded:
- Average run time under 3.5 minutes per sample
- Continuous calibration validation (CCV) recoveries within ±10% of target values for all analytes
- Relative standard deviations below 5% for repeated analysis of reference water
Benefits and practical applications
The combination of ORS, advanced ion optics and detector delivers:
- Lower detection limits and improved interference removal without reactive gases
- Simplified method setup under a universal condition set
- High throughput with run times below four minutes per sample
- Enhanced robustness for high-matrix samples via HMI integration
These advantages support routine drinking water monitoring, regulatory compliance testing and broader environmental or industrial analyses without instrument reconfiguration.
Future trends and applications
Further developments may include:
- Expansion of reaction cell gases or modes for even more challenging matrices
- Integration of real-time data correction algorithms leveraging universal collision mode data
- Application of the same streamlined methods to complex environmental samples such as soils, sediments and sludges
- Automation enhancements to support fully unattended multi-sample workflows
Conclusion
The Agilent 7700x/7800 ICP-MS with ORS in helium mode provides a robust, high-sensitivity solution for trace metal analysis in drinking waters. Universal interference removal, low detection limits and simplified operation yield reliable data quality and high productivity. The system’s wide dynamic range and HMI capability ensure consistent performance across a variety of sample matrices without the need for reactive cell gases or complex corrections.
Reference
Steve Wilbur. Agilent 7700x/7800 Series ICP-MS Octopole Reaction System for Drinking Water Analysis. Agilent Technologies Application Note 5990-4315EN, July 2015.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Agilent ICP-MS Journal (September 2009 – Issue 39)
2009|Agilent Technologies|Others
Agilent ICP-MS Journal September 2009 – Issue 39 7700 Series ICP-MS Special Issue Inside this Issue 2 Clearly Better ICP-MS: Introducing the New 7700 Series, Agilent 7700 Series Stars on YouTube 3 New ORS Insures More Effective Interference Removal in…
Key words
icp, icppolyatomic, polyatomicremoval, removalagilent, agilentmode, modeplasma, plasmacell, cellcertified, certifiedhmi, hmiinterferences, interferenceselement, elementmetals, metalsgas, gasmatrix, matrixdrinking
Agilent 7900 ICP-MS simplifies drinking water analysis
2014|Agilent Technologies|Applications
Agilent 7900 ICP-MS simplifies drinking water analysis Application note Environmental Authors Tetsushi Sakai Agilent Technologies, Japan Ed McCurdy Agilent Technologies, UK Introduction Ensuring the quality of drinking water is a primary goal for public health bodies around the world. Most…
Key words
drinking, drinkingwater, waterelements, elementspolyatomic, polyatomicregulated, regulatedmode, modequality, qualityconc, concicp, icpallowable, allowablegas, gasinterferences, interferencesmatrix, matrixregulations, regulationsomega
Simple, Reliable Analysis of High Matrix Samples According to US EPA Method 6020A ICP-MS
2021|Agilent Technologies|Applications
Application Note Environmental Simple, Reliable Analysis of High Matrix Samples According to US EPA Method 6020A ICP-MS All samples could be analyzed without prior knowledge of their composition Authors Steve Wilbur, Craig Jones Agilent Technologies, Inc. USA Introduction US EPA…
Key words
sediment, sedimentgas, gassoil, soilmatrix, matrixelement, elementsamples, samplescrm, crmriver, riverpolyatomic, polyatomicsame, samellccv, llccvhmi, hmiverification, verificationcontinuing, continuingcalibration
Investigations into the use of helium collision mode and aerosol dilution for ultra-trace analysis of metals in mineral reference materials
2013|Agilent Technologies|Applications
Investigations into the use of helium collision mode and aerosol dilution for ultra-trace analysis of metals in mineral reference materials White Paper Mining Author Kazuo Yamanaka1 Steve Wilbur2 Agilent Technologies, 1 Hachioji, Japan 2 Everett, WA, US Introduction The analytical…
Key words
ppb, ppbindi, indigas, gascert, certblk, blkelements, elementsconc, concanalysis, analysisore, oreicb, icbresult, resultmode, moderatio, ratioccb, ccbflow