Agilent ICP-MS Journal (September 2009 – Issue 39)
Others | 2009 | Agilent TechnologiesInstrumentation
The introduction of the Agilent 7700 Series ICP-MS represents a major advance in quadrupole ICP-MS design, directly addressing longstanding challenges in matrix tolerance, dynamic range and spectral interference removal. By combining a fast frequency-matching RF generator, a high-performance ion lens, the third-generation Octopole Reaction System (ORS3) and an integrated High-Matrix Introduction (HMI) interface, the 7700 Series delivers more accurate, robust and productive trace and major element analyses across a wide variety of sample types.
This collection of technical reports aimed to:
The 7700 Series incorporates multiple innovations to achieve its performance gains:
Interference Removal in Complex Matrices
The ORS3 performance was tested using a mixed acid/organic matrix known to generate multiple polyatomic overlaps. No-gas and H2 reaction modes left significant residual interferences and created new overlaps, whereas helium mode with ORS3 achieved complete removal of matrix-derived polyatomics without introducing artefacts.
Drinking Water Analysis
In a continuous 11-hour sequence of 190 samples including high-TDS mineral waters and EPA-required calibration checks, no continuing calibration verification failures occurred. Method detection limits ranged from sub-ppt levels (e.g., Hg 1.2 ppt, U 1.7 ppt) to low tens of ppt for first-row transition metals, and robust plasma settings were maintained via One-click Plasma Setting.
Food Sample Screening
Certified reference materials (fish protein, mussel tissue, whole egg powder) were digested by microwave and analyzed in a single helium cell-gas run. Results for trace metals (As, Se, Cd, Pb, etc.) agreed within certified uncertainties, demonstrating that the 7700x can replace multiple dedicated techniques (ICP-OES, GFAAS, CV-AAS) for rapid total metal screening.
The 7700 Series offers the following advantages for routine and research laboratories:
Upcoming developments and application areas include:
The Agilent 7700 Series achieves the long-promised performance of quadrupole ICP-MS by uniting innovative hardware and software advances. Its universal helium collision-cell method, extensive dynamic range and high matrix tolerance simplify method development, improve data quality and boost laboratory productivity across environmental, food, semiconductor and research applications.
ICP/MS
IndustriesEnvironmental, Food & Agriculture
ManufacturerAgilent Technologies
Summary
Significance of the Topic
The introduction of the Agilent 7700 Series ICP-MS represents a major advance in quadrupole ICP-MS design, directly addressing longstanding challenges in matrix tolerance, dynamic range and spectral interference removal. By combining a fast frequency-matching RF generator, a high-performance ion lens, the third-generation Octopole Reaction System (ORS3) and an integrated High-Matrix Introduction (HMI) interface, the 7700 Series delivers more accurate, robust and productive trace and major element analyses across a wide variety of sample types.
Objectives and Study Overview
This collection of technical reports aimed to:
- Describe the design features and key performance specifications of the new Agilent 7700 Series (7700x and 7700s).
- Demonstrate the enhanced interference removal capabilities of the ORS3 in helium collision mode for complex matrices.
- Evaluate the instrument’s analytical performance in routine applications, including drinking water and food sample analysis.
- Highlight the practical benefits of a universal, single-gas collision-cell method and streamlined software workflows.
Methodology and Instrumentation
The 7700 Series incorporates multiple innovations to achieve its performance gains:
- Fast, frequency-matching 27 MHz RF generator tolerant of high organic loads.
- Peltier-cooled spray chamber and low-flow sample introduction for improved stability.
- Redesigned off-axis ion lens delivering higher sensitivity and lower background.
- High-Matrix Introduction (HMI) system for direct analysis of samples up to several percent total dissolved solids.
- Third-generation Octopole Reaction System (ORS3) optimized for helium collision mode using kinetic energy discrimination.
- High-transmission, matrix-tolerant interface and high-performance vacuum system.
- Hyperbolic quadrupole mass analyzer coupled to a simultaneous dual-mode detector covering nine orders of dynamic range.
- Agilent MassHunter Software with One-click Plasma Setting, automated tuning and comprehensive data reporting.
Key Results and Discussion
Interference Removal in Complex Matrices
The ORS3 performance was tested using a mixed acid/organic matrix known to generate multiple polyatomic overlaps. No-gas and H2 reaction modes left significant residual interferences and created new overlaps, whereas helium mode with ORS3 achieved complete removal of matrix-derived polyatomics without introducing artefacts.
Drinking Water Analysis
In a continuous 11-hour sequence of 190 samples including high-TDS mineral waters and EPA-required calibration checks, no continuing calibration verification failures occurred. Method detection limits ranged from sub-ppt levels (e.g., Hg 1.2 ppt, U 1.7 ppt) to low tens of ppt for first-row transition metals, and robust plasma settings were maintained via One-click Plasma Setting.
Food Sample Screening
Certified reference materials (fish protein, mussel tissue, whole egg powder) were digested by microwave and analyzed in a single helium cell-gas run. Results for trace metals (As, Se, Cd, Pb, etc.) agreed within certified uncertainties, demonstrating that the 7700x can replace multiple dedicated techniques (ICP-OES, GFAAS, CV-AAS) for rapid total metal screening.
Benefits and Practical Applications
The 7700 Series offers the following advantages for routine and research laboratories:
- Universal interference removal under a single helium collision-cell mode eliminates method complexity and sample-specific corrections.
- Broad dynamic range (sub-ppt to >500 ppm) in a single acquisition simplifies quantitation across diverse analyte concentrations.
- High matrix tolerance extends analysis to challenging sample types without dilution or alternative techniques.
- Automated plasma optimization and tuning reduces setup time and operator variability.
- Compact footprint and reduced weight (30% smaller than the previous generation) save laboratory space.
Future Trends and Potential Applications
Upcoming developments and application areas include:
- Integration with hyphenated techniques (LC-ICP-MS, GC-ICP-MS) for speciation analysis of elements such as As, Se and Hg.
- Exploration of alternative cell gases and reaction chemistries for targeted removal of challenging interferences.
- Increased automation and remote monitoring capabilities for high-throughput and regulated environments.
- Application to emerging fields such as nanoparticle characterization, bioimaging and metallomics.
- Enhanced data analytics and cloud-based processing for improved laboratory informatics.
Conclusion
The Agilent 7700 Series achieves the long-promised performance of quadrupole ICP-MS by uniting innovative hardware and software advances. Its universal helium collision-cell method, extensive dynamic range and high matrix tolerance simplify method development, improve data quality and boost laboratory productivity across environmental, food, semiconductor and research applications.
References
- Agilent Technologies. The Agilent 7700 Series ICP-MS: Extraordinary Design. Unparalleled Performance. Brochure, 5990-4025EN.
- Agilent Technologies. The Agilent 7700 Advantage for Drinking Water Analysis. Application Note, 5990-4315EN.
- Agilent Technologies. Simple, Rapid Analysis of Trace Metals in Foods Using the Agilent 7700x ICP-MS. Application Note, 5990-4539EN.
- Agilent Technologies. Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS. Technical Overview, 5990-3236EN.
- Agilent Technologies. Agilent 7700 Series ICP-MS – Design, Technology and Principles of Operation. Wall Poster, 5990-4234EN.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Agilent ICP-MS Journal (May 2010 – Issue 42)
2010|Agilent Technologies|Others
Agilent ICP-MS Journal May 2010 – Issue 42 Inside this Issue Prizes s for Paperent's for Agil Abstracts eciation New Sp pt. Due by Se Handbook pple iPads to A 30, 2010. ee page 8 for be Won. S tails.…
Key words
icp, icpagilent, agilentpbdes, pbdesinterface, interfacedecabromodiphenyl, decabromodiphenylhplc, hplchigh, highpbde, pbdepolybrominated, polybrominatedhmi, hmiinjector, injectormatrix, matrixcollision, collisionplasma, plasmafeces
Agilent 7700 Series ICP-MS
2010|Agilent Technologies|Brochures and specifications
Agilent 7700 Series ICP-MS Extraordinary design. Unparalleled performance. The Agilent 7700 Series ICP-MS Advancing ICP-MS technology. Simplifying trace metals analysis. Whether you need to analyze hundreds of complex samples in less time, or confidently detect ultra trace metals in high…
Key words
icp, icppolyatomic, polyatomicenergy, energyyour, yourroutine, routineagilent, agilentcell, cellremoval, removalhyperbolic, hyperbolicrefitting, refittingions, ionshigh, highsts, stspolyatomics, polyatomicsever
Agilent ICP-MS Journal (March 2012 – Issue 49)
2012|Agilent Technologies|Others
Agilent ICP-MS Journal March 2012 – Issue 49 Inside this Issue 2-3 Agilent’s New 8800 Triple Quadrupole ICP-MS: Hardware and Technology Introduction 4-5 Agilent’s 8800 ICP Triple Quad: Modes of Operation 6 On-line Isotope Dilution Analysis of Selenium using the…
Key words
icp, icpions, ionscell, cellreaction, reactionmode, modeconventional, conventionalquadrupole, quadrupoleisotope, isotopeproduct, productmatrix, matrixagilent, agilentinterferences, interferencesnew, newhandbook, handbookion
Agilent ICP-MS Journal (February 2010 – Issue 41)
2010|Agilent Technologies|Others
Agilent ICP-MS Journal February 2010 – Issue 41 Inside this Issue 2 3 Improving Data Quality in ICP-MS with Qualifier Ions 4-5 High Throughput Analysis of Lead in Whole Blood using ICP-MS with ISIS-DS 6 User article: Simultaneous Speciation Analysis…
Key words
icp, icpisis, isisspeciation, speciationagilent, agilentblood, bloodpolyatomic, polyatomicsrm, srmisotopes, isotopeshandbook, handbookcadph, cadphwinter, winterinterferences, interferencesqualifier, qualifierpharma, pharmamatrix