Difference in Quantifiable Concentration Ranges of UV-Vis Spectrophotometer and Fluorescence Spectrophotometer
Applications | 2020 | ShimadzuInstrumentation
The accurate quantification of chemical species across wide concentration ranges underpins many applications in environmental monitoring, pharmaceutical analysis and quality control. Spectrophotometric techniques such as UV-Vis absorbance and fluorescence detection offer complementary advantages in sensitivity, linearity and dynamic range. Understanding their comparative performance for a model analyte like rhodamine B can guide method selection in analytical laboratories.
The primary goal was to evaluate and compare the lower limits of detection and quantitation, calibration linearity and practical sensitivity of a UV-Vis spectrophotometer versus a fluorescence spectrophotometer using rhodamine B solutions. Key aims included:
The experiment employed two Shimadzu instruments under defined conditions:
Absorbance spectra showed strong linearity (R2 = 0.9999) in the 0.31–5 µg/mL range but suffered noise and reduced linearity at concentrations below 0.31 µg/mL. Fluorescence measurements produced highly linear calibration down to 0.013 µg/mL (R2 = 0.9991) with minimal baseline noise, although inner-filter effects and reabsorption caused curvature above 0.125 µg/mL.
Lower limit of quantitation and detection:
The fluorescence method demonstrated over 400-fold higher sensitivity versus UV-Vis for rhodamine B. Calibration linearity across concentration ranges showed the UV-Vis system excelled at higher levels, while fluorescence detection offered superior performance in trace determinations.
By aligning technique selection with target concentration ranges, analysts can maximize sensitivity and data quality. UV-Vis absorbance remains robust for moderate to high analyte levels, whereas fluorescence detection is ideal for trace-level quantitation in fields such as environmental trace analysis, biomedical assays and dye monitoring in manufacturing.
Emerging developments include integrated microfluidic sampling for reduced reagent use, time-resolved and multiplexed fluorescence to further lower detection limits, and hybrid systems combining absorbance and fluorescence channels. Advances in detector technology, data-driven calibration models and portable instrumentation will expand real-time, on-site monitoring capabilities.
This comparative study highlights the complementary roles of UV-Vis and fluorescence spectrophotometry. Appropriate instrument choice based on concentration range enables reliable quantification of rhodamine B. Fluorescence methods achieve superior sensitivity at trace levels, while UV-Vis spectroscopy offers stable performance at higher analyte loads.
UV–VIS spectrophotometry
IndustriesManufacturerShimadzu
Summary
Significance of the Topic
The accurate quantification of chemical species across wide concentration ranges underpins many applications in environmental monitoring, pharmaceutical analysis and quality control. Spectrophotometric techniques such as UV-Vis absorbance and fluorescence detection offer complementary advantages in sensitivity, linearity and dynamic range. Understanding their comparative performance for a model analyte like rhodamine B can guide method selection in analytical laboratories.
Objectives and Overview of the Study
The primary goal was to evaluate and compare the lower limits of detection and quantitation, calibration linearity and practical sensitivity of a UV-Vis spectrophotometer versus a fluorescence spectrophotometer using rhodamine B solutions. Key aims included:
- Measuring absorbance and fluorescence intensity across standard solutions (0.003–5 µg/mL for UV-Vis; 0.013–5 µg/mL for fluorescence).
- Establishing calibration curves in high- and low-concentration regions.
- Determining limits of detection (LOD) and quantitation (LOQ) and comparing instrument performance.
Methodology and Instrumentation
The experiment employed two Shimadzu instruments under defined conditions:
- UV-Vis spectrophotometer (UV-2600i): Wavelength range 300–700 nm, slit width 1.0 nm, scan speed medium, sampling pitch 1.0 nm, light source switching at 340 nm. Rhodamine B standards (0.003–5 µg/mL) were measured at 544 nm.
- Fluorescence spectrophotometer (RF-6000): Excitation at 544 nm, emission scanned 540–700 nm, bandwidth Ex. 5.0 nm/Em. 5.0 nm, scan speed 600 nm/min, sampling pitch 1.0 nm. Fluorescence intensity at 577 nm was recorded for 0.013–5 µg/mL standards.
Main Results and Discussion
Absorbance spectra showed strong linearity (R2 = 0.9999) in the 0.31–5 µg/mL range but suffered noise and reduced linearity at concentrations below 0.31 µg/mL. Fluorescence measurements produced highly linear calibration down to 0.013 µg/mL (R2 = 0.9991) with minimal baseline noise, although inner-filter effects and reabsorption caused curvature above 0.125 µg/mL.
Lower limit of quantitation and detection:
- UV-2600i LOQ: 0.019 µg/mL; LOD: 0.0056 µg/mL.
- RF-6000 LOQ: 4.3 × 10–5 µg/mL; LOD: 1.3 × 10–5 µg/mL.
The fluorescence method demonstrated over 400-fold higher sensitivity versus UV-Vis for rhodamine B. Calibration linearity across concentration ranges showed the UV-Vis system excelled at higher levels, while fluorescence detection offered superior performance in trace determinations.
Benefits and Practical Applications of the Method
By aligning technique selection with target concentration ranges, analysts can maximize sensitivity and data quality. UV-Vis absorbance remains robust for moderate to high analyte levels, whereas fluorescence detection is ideal for trace-level quantitation in fields such as environmental trace analysis, biomedical assays and dye monitoring in manufacturing.
Future Trends and Possibilities
Emerging developments include integrated microfluidic sampling for reduced reagent use, time-resolved and multiplexed fluorescence to further lower detection limits, and hybrid systems combining absorbance and fluorescence channels. Advances in detector technology, data-driven calibration models and portable instrumentation will expand real-time, on-site monitoring capabilities.
Conclusion
This comparative study highlights the complementary roles of UV-Vis and fluorescence spectrophotometry. Appropriate instrument choice based on concentration range enables reliable quantification of rhodamine B. Fluorescence methods achieve superior sensitivity at trace levels, while UV-Vis spectroscopy offers stable performance at higher analyte loads.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Application Handbook Food, Beverages, Agriculture - Release 2
2012|Shimadzu|Guides
Application Handbook Food, Beverages, Agriculture Release 2 Food, Beverages and Agriculture Regarding food, water, beverages and agricultural cropland, the increasing world population is one of the biggest challenges of mankind. How can access be provided to sufficient and safe food…
Key words
news, newsanalysis, analysismeasurement, measurementfood, foodusing, usingsample, samplewithout, withoutwater, watermrm, mrmacid, acidwere, wereflowrate, flowratemethod, methodmin, mindrinking
Food Safety Solutions - Application Data Book
2015|Shimadzu|Guides
Food Safety Solutions Application Data Book Foreword Excellence in Science Dear Valued Customers, Shimadzu believes in total customer satisfaction. Every Shimadzu instrument sold is backed up by our strong after-sales support team. Be it service support, or application support, we…
Key words
area, areamrm, mrmhch, hchmethyl, methylesi, esipumps, pumpsratio, ratiomode, modeparathion, parathiongas, gastemperature, temperaturersd, rsdname, namecompound, compoundtest
Multifaceted Spectrophotometric Analysis of Vitamins and Vitamin-like Substances
2025|Shimadzu|Applications
UV-Visible Spectrophotometer, Spectrofluorophotometer, Fourier Transform Infrared Spectrophotometer, and Infrared Raman Microscope Application News Multifaceted Spectrophotometric Analysis of Vitamins and Vitamin-like Substances Kazuki Sobue User Benefits Ultraviolet-visible spectrophotometers can measure the light absorption properties, ranging from ultraviolet to visible light,…
Key words
vitamin, vitaminfluorescence, fluorescenceacid, acidvitamins, vitaminsspectrophotometer, spectrophotometerintensity, intensityethanol, ethanolinfrared, infraredlike, likebiotin, biotinfolic, folicaminobenzoic, aminobenzoicraman, ramananthranilic, anthranilicpurified
Analysis of Residual Catalysts in Pharmaceuticals Using ICP Atomic Emission Spectrometry
2010|Shimadzu|Applications
LAAN-C-XX-E007 Medicine Application Note No.6 (Medicine) Analysis of Residual Catalysts in Pharmaceuticals Using ICP Atomic Emission Spectrometry T. Taniguchi, S. Kawano, T. Yamaguchi, J. Nagata, T. Nishine 1. Introduction In addition to containing active ingredients, pharmaceuticals contain a number of…
Key words
emission, emissionicp, icpsubstance, substancetosufloxacin, tosufloxacintosilate, tosilatepharmaceutical, pharmaceuticalatomic, atomiccatalysts, catalystsplasma, plasmaaddition, additionelement, elementresidual, residualelements, elementspharmaceuticals, pharmaceuticalswavelength