Analysis of Waste Samples According to US EPA Method 6010D

Applications | 2020 | Agilent TechnologiesInstrumentation
ICP-OES
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Importance of the Topic

Environmental monitoring of solid waste such as soils, sediments, and sludges demands precise trace element analysis to protect human health and meet regulatory requirements.
Inductively coupled plasma optical emission spectrometry (ICP-OES) following US EPA Method 6010D offers a performance-based framework for quantifying 31 elements after microwave-assisted digestion.

Objectives and Study Overview

This application note demonstrates the use of the Agilent 5800 Vertical Dual View (VDV) ICP-OES with an AVS 7 switching valve to streamline EPA Method 6010D workflows.
The study showcases smart software tools for method development, quality assurance, and maintenance management, validated through analyses of reference materials and matrix spike tests.

Methodology and Instrumentation

The Agilent 5800 VDV ICP-OES was equipped with an AVS 7 seven-port switching valve and an SPS 4 autosampler for high throughput.
Soil samples (0.5 g) were predigested with HNO₃ and microwave-digested per EPA Method 3051A, then diluted to 25 mL.
Internal standards yttrium and lutetium (10 mg/L) corrected matrix effects and easily ionizable element interferences.
Key smart tools in Agilent ICP Expert software:
  • IntelliQuant Screening: semiquantitative profiling of up to 70 elements, interference detection, and optimal wavelength recommendations;
  • Outlier Conditional Formatting (OCF): comparison of multiple wavelengths per element to flag inconsistent results;
  • Early Maintenance Feedback (EMF): alerts for scheduled maintenance based on sample count.


Main Results and Discussion

Instrument detection limits (IDLs) and method detection limits (MDLs) met EPA criteria (e.g., Ag MDL 0.0006 mg/L in solution, 0.0293 mg/kg in sample).
Single-point calibrations (0–250 mg/L for majors, 0–5 mg/L for traces) exhibited RSE < 2%.
Quality control checks—including calibration blanks, verifications, interference checks, and low-level spikes—complied with EPA 6010D acceptance limits.
SRM 2711a analyses achieved recoveries within ±10% for all elements except Sb (expected low due to absence of HCl in digestion).
Matrix spike recoveries in silica sand ranged 95–117% across 32 elements, confirming method accuracy.
Six-hour stability testing (272 analyses) produced ICV recoveries within ±10% and RSD < 2% for all elements.

Benefits and Practical Applications

  • AVS 7 valve reduces torch maintenance and contamination, enhancing uptime.
  • Smart software tools minimize remeasurements by pre-identifying interferences and selecting optimal wavelengths.
  • OCF and IntelliQuant provide rapid QC visualization and troubleshooting in complex matrices.
  • EMF optimizes maintenance scheduling, reducing consumable costs and unexpected downtime.


Future Trends and Potential Applications

Integration of advanced data analytics and AI for real-time interference correction and automated method optimization.
Higher multiplexing and faster cycle times to address increasing environmental and industrial sample loads.
Enhanced detector designs for ultra-trace element sensitivity and expanded dynamic ranges.

Conclusion

The Agilent 5800 VDV ICP-OES with AVS 7 and ICP Expert smart software fulfills EPA Method 6010D requirements for robust, accurate multi-element analysis of waste matrices.
Smart QA/QC and maintenance features ensure reliable performance over extended runs, simplifying both method development and routine laboratory operation.

References

  1. U.S. EPA. Method 6010D (SW-846). Revision 5. Inductively Coupled Plasma-Atomic Emission Spectrometry. U.S. Environmental Protection Agency; 2015.
  2. U.S. EPA. Method 3051A (SW-846). Revision 1. Microwave Assisted Acid Digestion of Sediments, Sludges, and Oils. U.S. Environmental Protection Agency; 2007.
  3. Agilent Technologies. Agilent ICP Expert Software: Powerful Software with Smart Tools for ICP-OES. Publication 5994-1517EN.
  4. Agilent Technologies. Agilent IntelliQuant Screening: Smarter and Quicker Semiquantitative ICP-OES Analysis. Publication 5994-1518EN.
  5. Agilent Technologies. Reduce Costs and Boost Productivity with the Advanced Valve System (AVS) 6 or 7 Port Switching Valve System. Publication 5991-6863EN.
  6. Agilent Technologies. Agilent IntelliQuant Software: For Greater Sample Insight and Simplified Method Development. Publication 5994-1516EN.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Analysis of Solid Waste Samples per Chinese Method HJ 781
Analysis of Solid Waste Samples per Chinese Method HJ 781
2019|Agilent Technologies|Applications
Application Note Environmental Analysis of Solid Waste Samples per Chinese Method HJ 781 Agilent 5800 VDV ICP-OES provides smart tools to streamline method development and improve confidence in results Author Peter Riles Agilent Technologies, Inc. Introduction Environmental agencies around the…
Key words
fitted, fittedradial, radialaxial, axialintelliquant, intelliquantcorrection, correctionfact, factelement, elementwaste, wasteoutlier, outlierindustrial, industrialmaintenance, maintenancebackground, backgroundwavelength, wavelengthmdl, mdlformatting
Analysis of Soils, Sediments, and Sludges by ICP-OES per US EPA 6010D
Application Brief Environmental Analysis of Soils, Sediments, and Sludges by ICP-OES per US EPA 6010D Improved data quality with the use of smart tools Analyze sample digests with more certainty The US Environmental Protection Agency (EPA) 6010D method is widely…
Key words
flag, flagconfidentiality, confidentialityoutlier, outlierintelliquant, intelliquantemf, emffeedback, feedbacklabel, labelocf, ocfquality, qualityformatting, formattingconditional, conditionalwaste, wastetough, toughmaintenance, maintenanceanalyzing
Analysis of Metals in Waste Samples Using Agilent 5800 Vertical Dual View (VDV) ICP-OES Systems
Consumable workflow ordering guide Analysis of Metals in Waste Samples Using Agilent 5800 Vertical Dual View (VDV) ICP-OES Systems Consumable workflow ordering guide for EPA 6010D In support of its waste management program, the US Environmental Protection Agency (EPA) has…
Key words
chamber, chamberpass, passdouble, doublepump, pumpspray, spraycyclonic, cyclonictubing, tubingwhite, whiteblack, blackocf, ocfordering, orderingblue, bluesolubilized, solubilizedmylist, mylistvdv
Automating the Workflow for the Analysis of Soils by ICP-OES
Application Note Environmental Automating the Workflow for the Analysis of Soils by ICP-OES Productive analysis of high matrix samples using the Agilent 5800 ICP-OES with Advanced Dilution System Author Introduction Daniel McCarthy To manage the large sample loads that are…
Key words
autodilution, autodilutiondilution, dilutionsample, sampleicp, icpoes, oesprescriptive, prescriptiveelement, elementdilutions, dilutionssrms, srmsresult, resultmeasured, measureduptake, uptakestock, stockcalibration, calibrationsamples
Other projects
GCMS
LCMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike