Analysis of Solid Waste Samples per Chinese Method HJ 781

Applications | 2019 | Agilent TechnologiesInstrumentation
ICP-OES
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


Accurate determination of trace and major elements in solid waste is essential for environmental monitoring and pollution control. Regulatory agencies rely on standardized methods to assess contamination levels and ensure safe disposal or remediation of industrial wastes.

Aims and Study Overview


This study evaluates the performance of the Agilent 5800 Vertical Dual View ICP-OES fitted with an AVS 7 switching valve for the analysis of 22 elements in solid waste samples following Chinese national standard HJ 781-2015. The objectives were to streamline method development, identify and correct spectral interferences, and demonstrate reliable results for industrial sludge reference materials.

Methodology and Instrumentation


Solid waste samples were digested using a four acid microwave program combining nitric acid hydrochloric acid hydrofluoric acid and perchloric acid in two stages. The digests were brought to a final matrix of 5 percent HNO3 and 15 percent HCl. An Agilent 5800 VDV ICP-OES with AVS 7 seven port switching valve and SPS 4 autosampler provided rapid sample uptake and reduced carryover. Key software features included IntelliQuant for semiquantitative screening and interference identification Outlier Conditional Formatting for result validation and Early Maintenance Feedback for scheduling consumable replacement. Automated background correction was achieved using Fitted Background Correction and Fast Automated Curve fitting Technique models.

Main Results and Discussion


Method detection limits for all 22 elements were well below those specified in HJ 781-2015. Analysis of NIST SRM 2782 industrial sludge yielded certified element recoveries within ±10 percent. Semiquantitative IntelliQuant screening of additional elements delivered results within ±30 percent of expected concentrations. Spike recovery tests for non certified elements also fell within ±10 percent. A six hour stability test with a quality control solution measured every ten samples showed recoveries within ±10 percent and relative standard deviations below 2.2 percent, confirming excellent precision and robustness.

Benefits and Practical Applications of the Method


  • High throughput enabled by rapid sample switching and integrated AVS 7 valve
  • Reduced unplanned downtime through early maintenance alerts based on sample count
  • Enhanced accuracy via automatic interference detection and correction using IntelliQuant and background correction tools
  • Full compliance with Chinese HJ 781-2015 standard for environmental solid waste analysis
  • Consistent analytical performance demonstrated by long term QC stability

Future Trends and Potential Applications


Advances in spectroscopic data analytics and machine learning could further automate interference correction and predictive maintenance. Potential extensions include analysis of soils sediments and other complex matrices. Digital twin models may streamline method optimization and reduce calibration efforts.
  • Integration of real time data analytics for smart method development
  • Automation of sample handling and instrument tuning
  • Application to broader environmental and industrial monitoring programs

Conclusion


The Agilent 5800 VDV ICP-OES with AVS 7 switching valve and ICP Expert software provides a robust sensitive and reliable workflow for analyzing solid waste per HJ 781-2015. Automated interference management and maintenance feedback enhance confidence in results and reduce downtime for routine environmental laboratories.

References


1. Solid Waste Determination of 22 metal elements Inductively coupled plasma optical emission spectrometry HJ 781 Standardization Administration of the Peoples Republic of China 2016
2. Agilent IntelliQuant Software Agilent publication 5994-1516EN
3. Fitted Background Correction Rapid automated background correction Agilent publication 5991-4836EN
4. Fast Automated Curve fitting Technique spectral correction Agilent publication 5991-4837EN

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Fast and Robust Analysis of Various Types of Waters by ICP-OES Following Method HJ 776-2015
Application Note Environmental Fast and Robust Analysis of Various Types of Waters by ICP-OES Following Method HJ 776-2015 Using smart tools to facilitate the analytical performance of the Agilent 5800 VDV ICP-OES Authors Milos Ridesic, Agilent Technologies, Inc. Introduction Clean,…
Key words
mdl, mdlintelliquant, intelliquantviewing, viewingicp, icpmaintenance, maintenancecrms, crmsalerts, alertsexpert, expertldr, ldrbackground, backgroundsample, samplenebulizer, nebulizerpump, pumpsmart, smartmeasured
Analysis of DTPA Extracted Soils per Chinese Standard HJ 804-2016
Application Note Environmental Analysis of DTPA Extracted Soils per Chinese Standard HJ 804-2016 Simplify method development and improve confidence in results with the Agilent 5800 VDV ICP-OES and IntelliQuant Screening Author Introduction Michael Mavrogenis Agilent Technologies, Inc. Understanding the elemental…
Key words
intelliquant, intelliquantdtpa, dtpafitted, fittedaxial, axialsoil, soilelements, elementselement, elementuptake, uptakeradial, radialbackground, backgroundwavelength, wavelengthpump, pumpcorrection, correctioneight, eightrinse
Analysis of Waste Samples According to US EPA Method 6010D
Application Note Environmental Analysis of Waste Samples According to US EPA Method 6010D Using the smart features of the Agilent 5800 VDV ICP-OES to simplify method development and quality assure the results Authors Introduction Peter Riles Agilent Technologies, Inc. In…
Key words
lloq, lloqintelliquant, intelliquanticv, icvmaintenance, maintenanceconfidentiality, confidentialityicsa, icsaocf, ocfrecovery, recoveryicb, icbscreening, screeningccb, ccbcorrection, correctionspike, spikebackground, backgroundicp
Determination of Metals in Recycled Li-ion Battery Samples by ICP-OES
Application Note Energy and Chemicals Determination of Metals in Recycled Li-ion Battery Samples by ICP-OES Quantifying 18 metals in “black mass” battery materials using the Agilent 5800 ICP-OES Author Introduction Shuping Li Lithium-ion batteries (LIBs) have been the power source…
Key words
fitted, fittedaxial, axialblack, blackradial, radialelements, elementsspike, spikelib, libfbc, fbcbattery, batterybackground, backgroundcorrection, correctionwaste, wastemeasured, measuredrecycling, recyclingsamples
Other projects
GCMS
LCMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike