Best Practice for Nucleic Acid Thermal Stability Measurements Using the Cary 3500 UV-Vis Spectrophotometer

Technical notes | 2022 | Agilent TechnologiesInstrumentation
UV–VIS spectrophotometry
Industries
Manufacturer
Agilent Technologies

Summary

Importance of the Topic


UV visible based thermal melting analysis is critical for assessing nucleic acid stability and interactions. It uses temperature dependent absorbance at 260 nm to monitor strand denaturation. These measurements support nucleic acid quantification quality control and provide insight into sequence length pH and buffer effects on melting behavior.

Objectives and Study Overview


This paper describes best practices for thermal stability measurements using a rapid Peltier controlled UV visible system. It outlines steps to optimize sample preparation instrument setup and data acquisition to achieve precise melting temperature determinations and reliable denaturation profiles.

Instrumentation Used


  • Cary 3500 UV Vis spectrophotometer with Peltier temperature control options including compact cell multicell and multizone modules
  • Quartz cuvettes of various path lengths with or without in cuvette temperature probes
  • Temperature probes and stirrers for in cuvette thermal feedback and mixing

Methodology and Instrumentation


Sample absorbance is recorded at 260 nm while ramping temperature. Key steps include selecting UV transparent quartz cuvettes cleaning with alkaline solution optimizing buffer pH and salt content and using stoppered cells or oil overlays to prevent evaporation. Condensation is avoided by thermal equilibration and nitrogen purging. Degassing protocols and controlled stirring ensure bubble free measurements. The software controls analysis wavelength bandwidth averaging time ramp rate data interval and temperature feedback source from block sensor or in cuvette probe.

Key Results and Discussion


Optimized settings of 1 to 2 nanometer spectral bandwidth and two second averaging time improve signal to noise. Slower block based ramp rates below half degree per minute ensure thermal equilibrium while probe based rates up to forty degrees per minute allow rapid profiling. Data intervals of one degree or less capture melting transitions sharply. Proper probe placement at beam center and stirring speed near nine hundred rpm yield uniform temperature.

Benefits and Practical Applications


These protocols enable accurate melting temperature determination for nucleic acid quality control sequence design and interaction studies. Precise thermal profiles support analyses of polymer cloud point drug protein interactions and protein denaturation. The approach reduces artifacts from condensation evaporation or degassing promoting confident interpretation of melting behavior.

Future Trends and Applications


Advances in multizone control will enable simultaneous kinetic studies across different temperatures or buffers. Integration with high throughput sampling and real time analysis will expand use in drug screening and structural biology. Improved software automation and probe technologies will further enhance precision and throughput.

Conclusion


Following these best practices for sample handling instrument setup and parameter selection ensures reproducible and high fidelity thermal melt data. Proper cuvette choice cleaning evaporation control and optimized software settings produce reliable melting temperatures for a range of nucleic acid and biomolecular studies.

References


  1. Mergny JL Lacroix L Analysis of thermal melting curves Oligonucleotides 2003 13(6) 515 37
  2. Good NE et al Hydrogen Ion Buffers for Biological Research Biochemistry 1966 5(2) 467 77
  3. Good NE Izawa S Hydrogen Ion Buffers Methods Enzymol 1972 24 53 68

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Fast Determination of Thermal Melt Temperature of Double-Stranded Nucleic Acids by UV-Vis Spectroscopy
Application Note Biopharmaceuticals Fast Determination of Thermal Melt Temperature of Double-Stranded Nucleic Acids by UV-Vis Spectroscopy Reproducible melting temperature (Tm) data at fast ramp rates using the Agilent Cary 3500 UV-Vis Authors Wesam Alwan and Mathieu Rault Agilent Technologies, Inc.…
Key words
ramp, rampderivative, derivativetemperature, temperaturesmoothing, smoothingmelt, meltmelting, meltinggξc, gξccary, caryherring, herringsperm, spermstranded, strandedrate, rateworkstation, workstationnucleic, nucleicinterval
Characterization of Protein Thermal Stability Using UV-Vis Spectroscopy
Application Note Biopharmaceuticals Characterization of Protein Thermal Stability Using UV-Vis Spectroscopy Benefits of the Agilent Cary 3500 UV‑Vis spectrophotometer with a Peltier temperature‑controlled multicell in protein therapeutic research Authors Carlos Santos Martin La Trobe Institute for Molecular Science (LIMS), La…
Key words
protein, proteintemperature, temperaturespectrophotometer, spectrophotometerpeltier, peltierthermal, thermalabsorbance, absorbancedsba, dsbaderivative, derivativemeasure, measuresmoothing, smoothingstability, stabilityvis, vismulticell, multicellecdsba, ecdsbatrobe
Phase Diagram of the Ouzo Effect Using Temperature-Controlled UV-Vis Spectroscopy
Application Note Academic Phase Diagram of the Ouzo Effect Using Temperature-Controlled UV-Vis Spectroscopy Simultaneous temperature ramping experiments using the Agilent Cary 3500 UV-Vis and the multizone feature of the Multicell sampling module Authors Introduction Marc-André Gagnon and Claire Cislak Agilent…
Key words
temperature, temperaturemulticell, multicellanethole, anetholediagram, diagramsolubility, solubilitypeltier, peltierpastis, pastisalcohol, alcoholphase, phaseramping, rampingcuvettes, cuvettesexperiments, experimentsvis, visethanol, ethanolmodule
Kinetics of an Oscillating Reaction using Temperature-Controlled UV-Vis Spectroscopy
Application Note Academia Kinetics of an Oscillating Reaction using Temperature-Controlled UV-Vis Spectroscopy Characterizing the Briggs-Rauscher reaction at four temperatures simultaneously using an Agilent Cary 3500 UV-Vis Authors Marc-André Gagnon, Claire Cislak, and Wesam Alwan Agilent Technologies, Inc. Abstract The Briggs-Rauscher…
Key words
briggs, briggsrauscher, rauscheroscillation, oscillationreaction, reactionmulticell, multicellcolor, colortemperature, temperaturevis, viskinetic, kinetictemperatures, temperaturespeltier, peltieroscillating, oscillatingdecay, decayspectrophotometer, spectrophotometermultizone
Other projects
GCMS
LCMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike