Increasing accuracy and precision of uranium isotope ratios by Thermal Ionization Mass Spectrometry using 1013 Ω amplifier technology

Technical notes | 2023 | Thermo Fisher ScientificInstrumentation
ICP/MS
Industries
Energy & Chemicals
Manufacturer
Thermo Fisher Scientific

Summary

Importance of Topic


Precise determination of uranium isotope ratios is critical for nuclear safeguards, environmental monitoring, geochronology and industrial quality control. Measurements of minor isotopes such as 234U and 236U and analyses of very small sample loads demand enhanced sensitivity and reduced detector noise. Traditional thermal ionization mass spectrometry (TIMS) with 1011 Ω amplifiers often requires larger sample amounts and suffers from limited dynamic range and elevated background noise.

Objectives and Study Overview


This study evaluates the performance of a Thermo Scientific Triton XT TIMS equipped with 1013 Ω amplifier technology. The goals are to compare precision and accuracy of uranium isotope ratio measurements between 1011 Ω and 1013 Ω amplifiers and to demonstrate reliable analysis down to picogram-level sample sizes. Certified reference materials NBS U-010 and U-500 were used to benchmark the method.

Applied Instrumentation


  • Thermo Scientific Triton XT thermal ionization mass spectrometer
  • Zone-refined double rhenium filaments for sample loading
  • Faraday cup collectors configured with 1011 Ω and 1013 Ω amplifiers

Methodology


Samples containing 0.5–20 ng total uranium were loaded onto rhenium filaments. A total evaporation protocol was applied, continuously adjusting filament current to maintain a reproducible evaporation profile and to collect all isotopes throughout the run. Two experimental series were conducted:
  • Comparison of 234U/238U and 236U/238U ratios in 20 ng U-500 and U-010 using both 1011 Ω and 1013 Ω amplifiers.
  • Measurement of 235U/238U in NBS U-010 with 1013 Ω detection of 235U over sample sizes of 0.2 ng, 10 pg and 5 pg (238U on 1011 Ω).

Main Results and Discussion


Use of 1013 Ω amplifiers reduced noise and improved ratio precision by factors of 4–5 for 234U/238U (down to ~0.05% 1 RSD) and up to 10 for 236U/238U (~0.08% 1 RSD) in 20 ng loads. For 235U/238U in 20 ng samples, precision reached 0.018% 1 RSD. Even at a total 235U load of 5 pg, 1013 Ω detection yielded ~0.5% 1 RSD. All measured ratios agreed with certified reference values within stated uncertainties.

Benefits and Practical Applications


Enhanced amplification extends TIMS capabilities to sub-nanogram and picogram sample sizes while maintaining high precision. This enables:
  • Improved nuclear safeguards and forensic analyses
  • Reduced sample consumption in geochronology and environmental studies
  • Higher accuracy in tracer and QA/QC applications

Future Trends and Opportunities


Further development of high-resistance amplifiers and integration with automated evaporation control will expand application scope. Combining 1013 Ω technology with multi-collector ICP-MS, micro-sampling techniques and advanced data processing (including machine-learning-based profile optimization) could enable in-situ measurements and real-time monitoring with unprecedented sensitivity.

Conclusion


The adoption of 1013 Ω amplifier technology on a Triton XT TIMS platform significantly improves precision and accuracy of uranium isotope ratio measurements for small sample loads. This advance supports more sensitive and resource-efficient analyses across nuclear, environmental and industrial applications.

Reference


  • Richter et al. (2011) J. Anal. At. Spectrom. 26, 550–564.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Uranium isotopic analysis for the nuclear industry
Uranium isotopic analysis for the nuclear industry
2019|Thermo Fisher Scientific|Applications
APPLICATION BRIEF 30238 Uranium isotopic analysis for the nuclear industry Authors Introduction Nicholas S. Lloyd, Claudia Bouman and Johannes B. Schwieters, Thermo Fisher Scientific, Bremen, Germany This application note describes the analytical performance of the Thermo Scientific™ Neptune XT™ Multicollector…
Key words
neptune, neptuneuranium, uraniumcertified, certifiedicp, icpreference, referencefaraday, faradaymulticollector, multicollectortriton, tritontims, timsisotope, isotopecounting, countingnuclear, nuclearmean, meandeviation, deviationfrom
Thermo Scientific Triton XT TIMS
Thermo Scientific Triton XT TIMS
2018|Thermo Fisher Scientific|Brochures and specifications
Thermo Scientific Triton XT TIMS Triton XT TIMS A new edition of the market leading Thermo Scientific™ Triton™ Series TIMS, capturing the best of technology for high-precision isotope ratio analysis. 2 Thermo Fisher Scientific: pioneering multicollector technologies, • First with…
Key words
triton, tritontims, timsrpq, rpqisotope, isotopeamplifier, amplifiercup, cupfaraday, faradaymulticollector, multicollectorfilament, filamention, ionmeasurements, measurementsamplifiers, amplifierssem, semcdd, cddprecision
Uranium Isotopic Analysis for the Nuclear Industry Using the Thermo Scientific NEPTUNE Plus MC-ICP-MS
Nicholas S. Lloyd, Claudia Bouman and Johannes B. Schwieters, Thermo Fisher Scientific, Bremen, Germany Ap plica t ion Note 3 0 23 8 Uranium Isotopic Analysis for the Nuclear Industry Using the Thermo Scientific NEPTUNE Plus MC-ICP-MS Key Words Instrumentation…
Key words
neptune, neptuneuranium, uraniumplus, pluscertified, certifiedicp, icpmethodology, methodologyisotope, isotopethermo, thermotriton, tritontest, testisotopic, isotopicdynode, dynodescientific, scientificreference, referencesem
Highly-sensitive uranium isotopic analysis for nuclear safeguards
APPLICATION NOTE 30269 Highly-sensitive uranium isotopic analysis for nuclear safeguards Authors Introduction Nicholas S. Lloyd, Claudia Bouman, and Johannes B. Schwieters Thermo Fisher Scientific, Bremen, Germany This application note describes the analytical performance of the Thermo Scientific™ Neptune XT™ multicollector…
Key words
neptune, neptuneicp, icpuranium, uraniumjet, jetinterface, interfaceatomic, atomicnuclear, nucleardata, datauncertainty, uncertaintypropagated, propagatedvalues, valuesratio, ratiorichter, richtermeasured, measuredrpq
Other projects
GCMS
LCMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike