How can you reduce helium consumption on the Thermo Scientific FlashSmart Elemental Analyzer?

Brochures and specifications | 2016 | Thermo Fisher ScientificInstrumentation
Elemental Analysis
Industries
Manufacturer
Thermo Fisher Scientific

Summary

Significance of the Topic


The Thermo Scientific FlashSmart Elemental Analyzer addresses a critical need in modern analytical laboratories to perform accurate, reproducible elemental analyses of carbon, hydrogen, nitrogen, sulfur and oxygen while controlling operating costs. With global helium shortages and rising prices, optimizing gas consumption or switching to alternative gases has become essential for high-throughput environments in research, quality control and industrial analytics.

Goals and Study Overview


This application note evaluates strategies to minimize helium use in the FlashSmart Analyzer during active sample runs and idle periods. It outlines five operational modes—ranging from software-driven standby to full argon carrier replacement—and compares their impact on helium savings, stabilization times and recommended usage scenarios.

Methodology and Used Instrumentation


The study focuses on the Thermo Scientific FlashSmart Elemental Analyzer coupled with the EagerSmart Data Handling Software. Selected optimization methods include:
  • Auto Stand-By via software to reduce carrier/reference flows to <3 mL/min.
  • Electrovalve closure (EV3/EV4) during automatic leak tests.
  • Manual valve switching between helium and nitrogen at the gas inlet.
  • MultiValve Control (MVC) module automating carrier gas switching.
  • Dedicated argon kit replacing helium as carrier and reference gas.

Main Results and Discussion


Implementation of software standby alone achieves over 90% helium reduction within 30 minutes. Electrovalve control and MVC approaches deliver nearly 100% helium savings during leak tests and extended idle periods, with return-to-analysis stabilization times under 30 minutes. The argon kit eliminates helium consumption entirely, though with slightly higher stabilization requirements when reverting to helium analyses.

Benefits and Practical Applications


These strategies help laboratories:
  • Drastically lower consumable costs by minimizing helium usage.
  • Maintain analytical throughput with rapid standby-to-ready transitions.
  • Leverage existing software and hardware modules without compromising data quality.

Future Trends and Possibilities


Ongoing developments may include real-time gas optimization algorithms, broader adoption of argon or mixed-gas carriers, and integration of predictive maintenance to further reduce idle purge requirements. Advances in low-flow furnace designs and alternative detector technologies could further diminish dependence on scarce noble gases.

Conclusion


By combining software controls, electrovalve management and optional argon carrier kits, the FlashSmart Elemental Analyzer delivers flexible solutions for minimizing helium use without sacrificing analytical performance. Laboratories can adapt these methods to their workflow demands, significantly reducing operating costs and improving sustainability.

References


Thermo Fisher Scientific. FlashSmart Elemental Analyzer Application Note. SN42269-EN 0716.
Thermo Fisher Scientific. EagerSmart Data Handling Software Technical Note 42270.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Optimized Helium Consumption in the Thermo Scientific FlashSmart Elemental Analyzer
TECHNICAL NOTE Dr. Liliana Krotz, Dr. Francesco Leone and Dr. Guido Giazzi Thermo Fisher Scientific, Milan, Italy Key Words Argon, Carrier Gas, Consumption, Cost per Analysis, Elemental Analysis, Helium, Nitrogen, MVC Module Goal To illustrate different options for the reduction…
Key words
flashsmart, flashsmarteagersmart, eagersmarthelium, heliumstand, standmvc, mvcargon, argonnitrogen, nitrogenconsumption, consumptiongas, gascarrier, carrierfurnace, furnacetemperatures, temperatureselectrovalves, electrovalvesanalyzer, analyzermodule
Thermo Scientific FlashSmart Elemental Analyzer
Thermo Scientific FlashSmart Elemental Analyzer
2024|Thermo Fisher Scientific|Brochures and specifications
Thermo Scientific FlashSmart Elemental Analyzer A single CHNS/O, NC, N/Protein Analyzer handling all applications Superior elemental analysis that grows with your needs Lead the way in elemental analysis with a single instrument handling all applications. You and your team can…
Key words
flashsmart, flashsmartchns, chnsmvc, mvcncs, ncsanalyzer, analyzermultivalve, multivalvechn, chnamerican, americansoftware, softwareautomated, automatedauto, autooxytune, oxytuneconfiguration, configurationsoils, soilsfibers
High Productivity of Thermo Scientific FlashSmart CHNS/CHNS Analyzer through the MultiValve Control (MVC) Module
TECHNICAL NOTE Dr. Liliana Krotz, Dr. Francesco Leone and Dr. Guido Giazzi Thermo Fisher Scientific, Milan, Italy Key Words Accuracy, Automation, CHNS, Flash Combustion, Organic Chemistry, Productivity, Unattended Analysis Goal To demonstrate the performance of the the FlashSmart Elemental Analyzer…
Key words
sulfanilamide, sulfanilamidemethionine, methioninechns, chnsmvc, mvcunk, unkmultivalve, multivalveeagersmart, eagersmartreactor, reactorflashsmart, flashsmartstd, stdhelium, heliumcombustion, combustionmonth, monthleft, leftrsd
High Automation of Thermo Scientific FlashSmart CHNS/O Analyzer using the MultiValve Control (MVC) Module
TECHNICAL NOTE Dr. Liliana Krotz, Dr. Francesco Leone, Walter Galotta and Dr. Guido Giazzi Thermo Fisher Scientific, Milan, Italy Key Words Accuracy, Automation, CHNS, Flash Combustion, Organic Chemistry, Oxygen, Pyrolysis, Unattended Analysis Goal To demonstrate the performance of the MultiValve…
Key words
bbot, bbotsulphanilamide, sulphanilamidemethionine, methioninechns, chnsunk, unkmvc, mvcacceptable, acceptableeagersmart, eagersmartoxygen, oxygenrsd, rsdsulfanilamide, sulfanilamiderange, rangeflashsmart, flashsmartstd, stdhelium
Other projects
GCMS
LCMS
Follow us
FacebookLinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike