Moisture and Ash Determination in Flour (TGA801)
Applications | 2020 | LECOInstrumentation
Accurate measurement of moisture and ash in flour products is critical for maintaining quality safety and consistency in food production Moisture affects texture taste microbial stability and yield calculations while ash content serves as an indicator of mineral content and potential contaminants such as bran in refined flours
The primary goal of this application note is to demonstrate a reliable thermogravimetric protocol for simultaneous determination of moisture and ash in different flour matrices using the LECO TGA801 macro thermogravimetric analyzer The study adapts AOAC Official Methods 925.10 and 923.03 to the TGA801 instrument specifications and evaluates performance with corn wheat and barley flours
A representative sample of approximately one gram of flour is weighed into a large ceramic crucible Sample weight is maintained between 0.8 and 1.2 grams The thermogravimetric procedure comprises two sequential steps
1 Drying step ramp from ambient to 130 °C at 10 °C per minute followed by a 30 minute hold until mass constancy within 0.0005 grams
2 Ashing step ramp from 130 °C to 550 °C at 20 °C per minute followed by a 30 minute hold in air atmosphere Flow rate is maintained at ten liters per minute Total analysis time is approximately four hours for each batch of up to nineteen samples
Typical results for ten replicates of each flour type show high precision and reproducibility
This thermogravimetric method provides laboratories with an automated high throughput solution for flour quality control Alignment with AOAC official protocols ensures compliance in food safety and labeling assessments The ability to run up to nineteen samples concurrently reduces turnaround time and improves laboratory productivity
Advances in thermogravimetric instrumentation may include real time data analytics for process control integration with near infrared spectroscopy for complementary compositional analysis and miniaturized furnace designs for rapid on line monitoring Further development of method libraries can expand application to other food matrices and complex ingredient blends
The LECO TGA801 based protocol delivers accurate reproducible moisture and ash determinations in flour samples Method performance meets stringent AOAC criteria and offers significant efficiency gains for routine quality assurance in food laboratories
Thermal Analysis
IndustriesFood & Agriculture
ManufacturerLECO
Summary
Importance of the Topic
Accurate measurement of moisture and ash in flour products is critical for maintaining quality safety and consistency in food production Moisture affects texture taste microbial stability and yield calculations while ash content serves as an indicator of mineral content and potential contaminants such as bran in refined flours
Objectives and Study Overview
The primary goal of this application note is to demonstrate a reliable thermogravimetric protocol for simultaneous determination of moisture and ash in different flour matrices using the LECO TGA801 macro thermogravimetric analyzer The study adapts AOAC Official Methods 925.10 and 923.03 to the TGA801 instrument specifications and evaluates performance with corn wheat and barley flours
Methodology
A representative sample of approximately one gram of flour is weighed into a large ceramic crucible Sample weight is maintained between 0.8 and 1.2 grams The thermogravimetric procedure comprises two sequential steps
1 Drying step ramp from ambient to 130 °C at 10 °C per minute followed by a 30 minute hold until mass constancy within 0.0005 grams
2 Ashing step ramp from 130 °C to 550 °C at 20 °C per minute followed by a 30 minute hold in air atmosphere Flow rate is maintained at ten liters per minute Total analysis time is approximately four hours for each batch of up to nineteen samples
Used Instrumentation
- LECO TGA801 macro thermogravimetric analyzer with dual furnace capability allowing parallel measurement of multiple samples
- Ceramic crucibles with mass range nineteen to thirty grams and lid assembly
- Double ended scoop for consistent sample transfer
Main Results and Discussion
Typical results for ten replicates of each flour type show high precision and reproducibility
- Corn flour average moisture seven point 39 percent standard deviation zero point 02 percent ash zero point 44 percent SD zero point 03 percent dry ash zero point 47 percent SD zero point 03 percent
- Wheat flour average moisture seven point 52 percent SD zero point 02 percent ash one point 49 percent SD zero point 07 percent dry ash one point 62 percent SD zero point 07 percent
- Barley flour average moisture seven point 92 percent SD zero point 03 percent ash one point 86 percent SD zero point 04 percent dry ash two point 02 percent SD zero point 04 percent
Benefits and Practical Applications
This thermogravimetric method provides laboratories with an automated high throughput solution for flour quality control Alignment with AOAC official protocols ensures compliance in food safety and labeling assessments The ability to run up to nineteen samples concurrently reduces turnaround time and improves laboratory productivity
Future Trends and Applications
Advances in thermogravimetric instrumentation may include real time data analytics for process control integration with near infrared spectroscopy for complementary compositional analysis and miniaturized furnace designs for rapid on line monitoring Further development of method libraries can expand application to other food matrices and complex ingredient blends
Conclusion
The LECO TGA801 based protocol delivers accurate reproducible moisture and ash determinations in flour samples Method performance meets stringent AOAC criteria and offers significant efficiency gains for routine quality assurance in food laboratories
References
- AOAC Official Method 925.10 Solids and Moisture in Flour
- AOAC Official Method 923.03 Ash of Flour
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Moisture and Ash Determination in Animal Feed (TGA801)
2020|LECO|Applications
Application Note Instrument: TGA801 EMPOWERING RESULTS Moisture and Ash Determination in Animal Feed LECO Corporation; Saint Joseph, Michigan USA Introduction An accurate determination of moisture and ash content in feed products provides important information related to the food quality and…
Key words
preset, presetash, ashconstancy, constancycrucible, cruciblemoisture, moisturestep, steptype, typecalculation, calculationweight, weightdisabled, disabledmass, massavg, avglids, lidsequation, equationmethod
Determination of Moisture, Volatile, and Ash in Biomass/Plant Tissue (TGA801)
2021|LECO|Applications
Application Note Instrument: TGA801 EMPOWERING RESULTS Determination of Moisture, Volatile, and Ash in Biomass/Plant Tissue LECO Corporation; Saint Joseph, Michigan USA Introduction Method Step Parameters - Moisture Moisture, volatile matter, and ash are important constituents for the quality characterization of…
Key words
preset, presetash, ashstep, stepconstancy, constancycrucible, cruciblecalculation, calculationtype, typemoisture, moisturedry, dryvolatile, volatiledisabled, disabledmass, massmethod, methodcalculations, calculationslids
Determination of Moisture and Ash in Graphite (TGA801)
2023|LECO|Applications
Application Note Instrument: TGA801 EMPOWERING RESULTS Determination of Moisture and Ash in Graphite LECO Corporation; Saint Joseph, Michigan USA Introduction Method Reference* Graphite is a stable form of carbon that exists in nature or can be synthetically produced. Graphite has…
Key words
ash, ashconstancy, constancypreset, presetstep, stepcrucible, cruciblemoisture, moisturegraphite, graphitetype, typeavg, avglids, lidsweight, weightcalculation, calculationatmosphere, atmospheremass, massactive
Determination of Moisture and Ash in Plant Tissue (TGA801)
2020|LECO|Applications
Application Note Instrument: TGA801 EMPOWERING RESULTS Determination of Moisture and Ash in Plant Tissue LECO Corporation; Saint Joseph, Michigan USA Introduction General Method Parameters The moisture content of plant tissue is often used to monitor the drying and curing process…
Key words
ash, ashmoisture, moisturepreset, presetconstancy, constancycrucible, cruciblestep, steptype, typecalculation, calculationplant, plantweight, weightmass, masstissue, tissuedisabled, disabledmethod, methodavg