Understanding Calibration for Glow Discharge Atomic Emission Spectrometry (GD-AES)
Technical notes | 2011 | LECOInstrumentation
Accurate calibration in Glow Discharge Atomic Emission Spectrometry (GD-AES) underpins reliable elemental analysis in industries ranging from metallurgy to surface coatings. Ensuring traceability to certified reference materials and maintaining linear response over broad concentration ranges are critical for quality control, materials development, and failure analysis.
The document reviews general considerations for GD-AES calibration on the LECO GDS850A instrument. It addresses both bulk quantitative analysis and quantitative depth profiling (QDP), outlining approaches to establish robust working curves, correct for sputter rate variability, and apply calibrations across diverse material matrices.
Sampling relies on a Grimm-style glow discharge lamp with defined anode diameters (2 mm and 4 mm) and argon sputtering under controlled partial pressure. Calibration curves are generated by measuring certified reference materials traceable to NIST or ISO Guide 31 standards at multiple specimen locations.
Bulk calibration yields linear working curves spanning ppm to high-percent concentration levels, enabling accurate interpolation and modest extrapolation. Figure 1 (Pb in Cu) demonstrates normalized linear response up to >16 % concentration. QDP calibration uses sputter rate correction to unify working curves across different alloys. Without correction, calibration families cluster by matrix; with correction, a single linear curve is recovered (Figures 4 and 5). Depth profiles (Figure 6) illustrate transition from Zn coating to Fe substrate, quantifiable without external thickness standards.
Advancements in detector technology and data processing will further improve sensitivity and dynamic range. Integration of automated sputter rate determination and machine-learning calibration algorithms may streamline method development. Expanding certified reference material libraries for novel alloy systems will broaden GD-AES adoption in emerging materials research.
GD-AES calibration on the LECO GDS850A platform demonstrates reliable, traceable elemental quantification for both bulk and depth-resolved analyses. Sputter rate correction is essential for multi-matrix calibrations, enabling unified working curves and accurate profiling of layered structures without auxiliary thickness measurements.
GD/MP/ICP-AES, Elemental Analysis
IndustriesMaterials Testing, Energy & Chemicals
ManufacturerLECO
Summary
Importance of the Topic
Accurate calibration in Glow Discharge Atomic Emission Spectrometry (GD-AES) underpins reliable elemental analysis in industries ranging from metallurgy to surface coatings. Ensuring traceability to certified reference materials and maintaining linear response over broad concentration ranges are critical for quality control, materials development, and failure analysis.
Aims and Overview of the Study
The document reviews general considerations for GD-AES calibration on the LECO GDS850A instrument. It addresses both bulk quantitative analysis and quantitative depth profiling (QDP), outlining approaches to establish robust working curves, correct for sputter rate variability, and apply calibrations across diverse material matrices.
Methodology and Instrumentation
Sampling relies on a Grimm-style glow discharge lamp with defined anode diameters (2 mm and 4 mm) and argon sputtering under controlled partial pressure. Calibration curves are generated by measuring certified reference materials traceable to NIST or ISO Guide 31 standards at multiple specimen locations.
Used Instrumentation
- LECO GDS850A Glow Discharge AES system
- Grimm-style glow discharge lamp (7 mm or 15 mm o-ring configurations)
- Array detectors and photomultiplier tubes for signal acquisition
Main Results and Discussion
Bulk calibration yields linear working curves spanning ppm to high-percent concentration levels, enabling accurate interpolation and modest extrapolation. Figure 1 (Pb in Cu) demonstrates normalized linear response up to >16 % concentration. QDP calibration uses sputter rate correction to unify working curves across different alloys. Without correction, calibration families cluster by matrix; with correction, a single linear curve is recovered (Figures 4 and 5). Depth profiles (Figure 6) illustrate transition from Zn coating to Fe substrate, quantifiable without external thickness standards.
Benefits and Practical Applications
- Robust quantification for bulk metals and thin films
- Elimination of profilometers for depth profiling
- Applicability to galvanizing, electroplating, PVD/CVD, heat treatments, oxides, and painted layers
- Traceability to recognized standards for regulatory compliance
Future Trends and Potential Applications
Advancements in detector technology and data processing will further improve sensitivity and dynamic range. Integration of automated sputter rate determination and machine-learning calibration algorithms may streamline method development. Expanding certified reference material libraries for novel alloy systems will broaden GD-AES adoption in emerging materials research.
Conclusion
GD-AES calibration on the LECO GDS850A platform demonstrates reliable, traceable elemental quantification for both bulk and depth-resolved analyses. Sputter rate correction is essential for multi-matrix calibrations, enabling unified working curves and accurate profiling of layered structures without auxiliary thickness measurements.
Reference
- ASTM E135 – Terminology for Analytical Chemistry for Metals, Ores, and Related Materials
- ASTM E158 – Fundamental Calculations to Convert Intensities into Concentrations in Optical Emission Spectrochemical Analysis
- ASTM E305 – Establishing and Controlling Spectrochemical Analytical Curves
- ASTM E406 – Using Controlled Atmospheres in Spectrochemical Analysis
- ASTM E415 – Optical Emission Vacuum Spectrometric Analysis of Carbon and Low-Alloy Steel
- ASTM E520 – Describing Photomultiplier Detectors in Emission and Absorption Spectrometry
- ASTM E1009 – Evaluating an Optical Emission Vacuum Spectrometer to Analyze Carbon and Low-Alloy Steel
- ASTM E1507 – Describing and Specifying the Spectrometer of an Optical Emission Direct Reading Instrument
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Quantitative Depth Profile (QDP) Analysis of Plated Samples
2006|LECO|Technical notes
Spectroscopy Performance Note Quantitative Depth Profile (QDP) Analysis of Plated Samples • Plating depths and coating weights • Composition of the coating and substrate • Surface treatments Improved corrosion resistance, increased wear resistance, or a decorative finish can be achieved…
Key words
micrometers, micrometersdepth, depthplating, platingpercent, percentgds, gdselectroless, electrolesscathode, cathodeglow, glownickel, nickelresurfacing, resurfacingargon, argonanode, anodeanalyte, analytenanometers, nanometersfewer
QDP Analysis of Galvanized Steel
2007|LECO|Technical notes
Spectroscopy Performance Note QDP Analysis of Galvanized Steel • Galvanizing Thickness and Coating Weight • Composition on the Coating and Substrate • Surface Treatments The application of zinc and zinc-alloy coatings to steel provides both galvanic and barrier corrosion protection.…
Key words
micrometers, micrometersdepth, depthpercent, percentweight, weightcoating, coatinggds, gdsanalyte, analyteglow, glowresurfacing, resurfacingargon, argoncoatings, coatingsgalvanized, galvanizedqdp, qdpzinc, zincfewer
Using Bulk Analysis to Characterize the Carbon Gradient in Carburized Steel (GDS900)
2020|LECO|Applications
Application Note Instrument: GDS900 EMPOWERING RESULTS Using Bulk Analysis to Characterize the Carbon Gradient in Carburized Steel LECO Corporation; Saint Joseph, Michigan USA Key Words: Surface Carbon Analysis, Sub-Surface Analysis, Case Analysis by Grinding, Compositional Analysis, Glow Discharge AES, Glow…
Key words
carburization, carburizationcarburized, carburizedsurface, surfacecarbon, carbonglow, glowbulk, bulkalloy, alloysteel, steeldischarge, dischargedropping, droppinggrinding, grindingburn, burnconcentration, concentrationanalyze, analyzecalibration
Preparation of Specimens for Bulk Analysis Using LECO Glow Discharge Spectrometers
2012|LECO|Technical notes
® Preparation of Specimens for Bulk Analysis Using LECO Glow Discharge Spectrometers Introduction Information in this document pertains to all LECO Glow Discharge Spectrometers, including the GDS850A and the GDS500A. The purpose of this material is to provide recommendations and…
Key words
leco, lecoglow, glowdischarge, dischargelamp, lampsample, samplesurface, surfacesamples, samplesanode, anodeanalyzed, analyzeddiallyl, diallylmetallographic, metallographicface, facespectroscopy, spectroscopymetal, metalring