Determination of Nitrogen/Protein in Cheese
Applications | 2023 | LECOInstrumentation
The accurate measurement of protein content in cheese is essential for nutritional labeling, quality control, and economic valuation. Rapid and precise determination supports consistent product quality and regulatory compliance in dairy processing.
This application note evaluates the LECO FP828 combustion (Dumas) nitrogen/protein determinator for cheese analysis. The goals are to demonstrate method speed, precision, and flexibility using different carrier gases (helium or argon) and aliquot dose loops (3 cm3 or 10 cm3), and to compare results across common cheese varieties.
Samples of uniform consistency are weighed into tin foil cups (~0.25 g) and analyzed without chemical digestion. The sequence includes system conditioning, blank determination, calibration/drift correction using reference materials, and sample analysis. Key procedural steps:
The LECO FP828 employs a vertical quartz furnace in a pure oxygen environment (950 °C), followed by an afterburner at 850 °C. Moisture is removed by a thermoelectric cooler before gases mix in a ballast. A selectable 3 cm3 or 10 cm3 aliquot dose loop delivers the sample to a thermal conductivity detector. Carrier gas options are helium (highest sensitivity) or argon.
Typical analyses on Parmesan, Swiss, Pepperjack, and Sharp Cheddar show:
The automated Dumas method on the FP828 provides:
Advancements may include integration with inline sampling, further miniaturization, alternative inert gases, and enhanced software automation. Expansion of Dumas applications to other dairy products and real-time process monitoring are promising areas.
The LECO FP828 offers a robust, fast, and safe method for nitrogen/protein determination in cheese. Its adaptability and reliability make it well-suited for high-throughput dairy laboratories and quality assurance programs.
Elemental Analysis
IndustriesFood & Agriculture
ManufacturerLECO
Summary
Importance of the Topic
The accurate measurement of protein content in cheese is essential for nutritional labeling, quality control, and economic valuation. Rapid and precise determination supports consistent product quality and regulatory compliance in dairy processing.
Study Objectives and Overview
This application note evaluates the LECO FP828 combustion (Dumas) nitrogen/protein determinator for cheese analysis. The goals are to demonstrate method speed, precision, and flexibility using different carrier gases (helium or argon) and aliquot dose loops (3 cm3 or 10 cm3), and to compare results across common cheese varieties.
Methodology
Samples of uniform consistency are weighed into tin foil cups (~0.25 g) and analyzed without chemical digestion. The sequence includes system conditioning, blank determination, calibration/drift correction using reference materials, and sample analysis. Key procedural steps:
- Condition the system with reference or sample replicates.
- Establish blank level from multiple blank runs (target SD ≤ 10 ppm for helium, ≤ 50 ppm for argon).
- Calibrate or drift-correct with certified reference material (e.g., EDTA, ~9.58 % N).
- Analyze cheese samples in replicates, recording nitrogen content.
Used Instrumentation
The LECO FP828 employs a vertical quartz furnace in a pure oxygen environment (950 °C), followed by an afterburner at 850 °C. Moisture is removed by a thermoelectric cooler before gases mix in a ballast. A selectable 3 cm3 or 10 cm3 aliquot dose loop delivers the sample to a thermal conductivity detector. Carrier gas options are helium (highest sensitivity) or argon.
Main Results and Discussion
Typical analyses on Parmesan, Swiss, Pepperjack, and Sharp Cheddar show:
- Consistent nitrogen values with standard deviations ≤ 0.02 % (helium) and ≤ 0.04 % (argon).
- Protein contents (using factor 6.38) closely match expected compositions (e.g., 30.6 % in Parmesan, 19.4 % in Swiss).
- Minimal impact of loop size on precision; helium yields slightly better sensitivity at low nitrogen levels.
Benefits and Practical Applications
The automated Dumas method on the FP828 provides:
- Analysis in under five minutes per sample without hazardous reagents.
- High precision and accuracy across cheese varieties.
- Operational flexibility via choice of carrier gas and aliquot loop size.
- Reduced reagent cost and simplified workflow for routine quality control.
Future Trends and Potential Applications
Advancements may include integration with inline sampling, further miniaturization, alternative inert gases, and enhanced software automation. Expansion of Dumas applications to other dairy products and real-time process monitoring are promising areas.
Conclusion
The LECO FP828 offers a robust, fast, and safe method for nitrogen/protein determination in cheese. Its adaptability and reliability make it well-suited for high-throughput dairy laboratories and quality assurance programs.
Reference
- ISO 14891: Milk and Milk Products — Determination of Nitrogen Content (Routine combustion method by Dumas principle), modified version used for data generation.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Determination of Nitrogen/Protein in Cheese (FP928)
2023|LECO|Applications
Application Note Instrument: FP928 EMPOWERING RESULTS Elemental Analysis | GC Mass Spectrometry | Metallography Determination of Nitrogen/Protein in Cheese LECO Corporation; Saint Joseph, Michigan USA Introduction Protein is one of the most significant nutrient components in cheese. The accurate and…
Key words
avg, avgcheese, cheeseprotein, proteincombustion, combustionyes, yesnitrogen, nitrogenconductivity, conductivityargon, argonthermal, thermaldumas, dumasaliquot, aliquotmass, massdeterminator, determinatorloop, loopballast
Determination of Nitrogen/Protein in Meat and Plant-Based Meat Alternatives
2023|LECO|Applications
Application Note EMPOWERING RESULTS Instrument: FP828 Elemental Analysis | GC Mass Spectrometry | Metallography Determination of Nitrogen/Protein in Meat and Plant-Based Meat Alternatives LECO Corporation; Saint Joseph, Michigan USA Introduction Protein is one of the most significant nutrient components in…
Key words
avg, avgdeli, delimeat, meatprotein, proteinentrapped, entrappedtwisting, twistingfoil, foilconductivity, conductivityargon, argonthermal, thermalballast, ballastyes, yesedges, edgesmass, massburn
Determination of Nitrogen/Protein in Whey and Powdered Dairy Products
2023|LECO|Applications
Application Note Instrument: FP828 EMPOWERING RESULTS Elemental Analysis | GC Mass Spectrometry | Metallography Determination of Nitrogen/Protein in Whey and Powdered Dairy Products LECO Corporation; Saint Joseph, Michigan USA Introduction Two of the main types of proteins within milk are…
Key words
avg, avgprotein, proteindairy, dairypowdered, powderedargon, argonwhey, wheymass, massburn, burnhelium, heliumthermal, thermalnitrogen, nitrogenappropriate, appropriateyes, yesconductivity, conductivityaliquot
Determination of Nitrogen/Protein in Soybean Meal
2022|LECO|Applications
Application Note Instrument: FP828 Series Determination of Nitrogen/Protein in Soybean Meal EMPOWERING RESULTS LECO Corporation; Saint Joseph, Michigan USA Introduction Soybean meal is the primary source of protein for the feed industry worldwide, and is an important source of protein…
Key words
avg, avgprotein, proteinmeal, mealsoybean, soybeannitrogen, nitrogenconductivity, conductivityaliquot, aliquotatmosphere, atmosphereballast, ballastargon, argonloop, loopburn, burnthermal, thermaltwisting, twistingmass