IR-Spectroscopic Analysis of Polymer Fillers and Compatibilizers
Applications | 2025 | Bruker OpticsInstrumentation
Modern plastics often contain multiple constituents—polymers, fillers, plasticizers and compatibilizers—that are blended to achieve tailored mechanical and chemical properties. Reliable identification and quantification of each component is essential for quality control and material performance assessment. Fourier transform infrared (FT-IR) spectroscopy, especially with Attenuated Total Reflection (ATR), offers rapid, accurate analysis without extensive sample preparation, making it a valuable tool in both research and industrial settings.
This application note demonstrates how a compact FT-IR system equipped with a diamond ATR module can identify and quantify fillers (talcum) and compatibilizer monomer content (styrene in SEBS copolymers). The aim is to illustrate method setup, calibration and validation workflows for routine polymer analysis.
Measurements were performed on a Bruker Optics ALPHA II FT-IR spectrometer fitted with a high-pressure platinum diamond ATR module. Key features:
1. Talcum identification in polypropylene samples:
• Pure PP and PP containing 10%, 20%, and 40% talcum were measured.
• Characteristic Si–O stretching bands at ~1000 and ~670 cm–1 and bending at ~400 cm–1 increase proportionally with talc content.
• Band intensities enable straightforward qualitative and quantitative assessment.
2. Styrene quantification in SEBS-type copolymers:
• Reference materials with known styrene contents (0%, 4.9%, 14.5%) and an unknown sample were analyzed.
• Two distinct fingerprint bands—aromatic C=C at 1492 cm–1 and ring deformation at 699 cm–1—were evaluated, with the 699 cm–1 band chosen for calibration.
• A three-point calibration in OPUS produced a linear fit (R2≈0.9999), allowing determination of the unknown sample at 3.6±0.1% styrene.
• Rapid (<1 min) verification of product specifications and batch consistency
• Non-destructive, minimal or no consumables required
• Applicability to solids, liquids, powders and laminates for surface-sensitive analysis
• User-friendly workflow suitable for operators with limited spectroscopy experience
Integration of FT-IR ATR with automated sampling systems and advanced chemometric algorithms will further enhance throughput and quantification accuracy. Emerging applications include in-line monitoring of extrusion processes, real-time compositional mapping of multilayer films and coupling with imaging modalities for spatially resolved analysis.
The compact ALPHA II FT-IR spectrometer with diamond ATR proves to be a robust, efficient solution for polymer filler and compatibilizer analysis. Its ease of use, speed and quantitative performance support routine quality control and research tasks across academic and industrial laboratories.
Bruker Optics Application Note M104 – IR-Spectroscopic Analysis of Polymer Fillers and Compatibilizers
FTIR Spectroscopy
IndustriesEnergy & Chemicals
ManufacturerBruker
Summary
Importance of IR Spectroscopy for Polymer Analysis
Modern plastics often contain multiple constituents—polymers, fillers, plasticizers and compatibilizers—that are blended to achieve tailored mechanical and chemical properties. Reliable identification and quantification of each component is essential for quality control and material performance assessment. Fourier transform infrared (FT-IR) spectroscopy, especially with Attenuated Total Reflection (ATR), offers rapid, accurate analysis without extensive sample preparation, making it a valuable tool in both research and industrial settings.
Study Objectives and Overview
This application note demonstrates how a compact FT-IR system equipped with a diamond ATR module can identify and quantify fillers (talcum) and compatibilizer monomer content (styrene in SEBS copolymers). The aim is to illustrate method setup, calibration and validation workflows for routine polymer analysis.
Methodology and Instrumentation
Measurements were performed on a Bruker Optics ALPHA II FT-IR spectrometer fitted with a high-pressure platinum diamond ATR module. Key features:
- Diamond crystal brazed in tungsten carbide for durability under high pressure
- Optional germanium crystal for high-absorbance or thin samples
- 360° rotatable pressure applicator for easy sample access
- Intuitive OPUS software with guided method creation and quantification tools
Main Results and Discussion
1. Talcum identification in polypropylene samples:
• Pure PP and PP containing 10%, 20%, and 40% talcum were measured.
• Characteristic Si–O stretching bands at ~1000 and ~670 cm–1 and bending at ~400 cm–1 increase proportionally with talc content.
• Band intensities enable straightforward qualitative and quantitative assessment.
2. Styrene quantification in SEBS-type copolymers:
• Reference materials with known styrene contents (0%, 4.9%, 14.5%) and an unknown sample were analyzed.
• Two distinct fingerprint bands—aromatic C=C at 1492 cm–1 and ring deformation at 699 cm–1—were evaluated, with the 699 cm–1 band chosen for calibration.
• A three-point calibration in OPUS produced a linear fit (R2≈0.9999), allowing determination of the unknown sample at 3.6±0.1% styrene.
Benefits and Practical Applications
• Rapid (<1 min) verification of product specifications and batch consistency
• Non-destructive, minimal or no consumables required
• Applicability to solids, liquids, powders and laminates for surface-sensitive analysis
• User-friendly workflow suitable for operators with limited spectroscopy experience
Future Trends and Potential Applications
Integration of FT-IR ATR with automated sampling systems and advanced chemometric algorithms will further enhance throughput and quantification accuracy. Emerging applications include in-line monitoring of extrusion processes, real-time compositional mapping of multilayer films and coupling with imaging modalities for spatially resolved analysis.
Conclusion
The compact ALPHA II FT-IR spectrometer with diamond ATR proves to be a robust, efficient solution for polymer filler and compatibilizer analysis. Its ease of use, speed and quantitative performance support routine quality control and research tasks across academic and industrial laboratories.
Reference
Bruker Optics Application Note M104 – IR-Spectroscopic Analysis of Polymer Fillers and Compatibilizers
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Analysis of Contaminations on a Plastic Part using the FT-IR Microscope LUMOS II
2021|Bruker|Applications
Application Note AN M101 Analysis of Contaminations on a Plastic Part using the FT-IR Microscope LUMOS II Introduction Contaminations of polymer products are often microscopically small and not easy to analyze. Without knowing the chemical composition of the defect, the…
Key words
lumos, lumosmicroscope, microscopeatr, atrstreaks, streakslateral, lateralpolycarbonate, polycarbonatecontaminations, contaminationsmotorized, motorizedfully, fullyblack, blackideal, idealwizard, wizardbruker, brukeralone, aloneobjective
Gathering Evidence by IR and Raman Spectroscopy
2015|Bruker|Brochures and specifications
Forensic Analysis Gathering Evidence by IR and Raman Spectroscopy Innovation with Integrity F T-IR Forensic Sciences apply Infrared and Raman spectroscopic methods for many years to analyze all types of unknown samples. These methods are proven to reveal the chemical…
Key words
raman, ramanvarnish, varnishfibers, fibersforgery, forgeryalpha, alphaforensic, forensicspectroscopy, spectroscopyhighs, highssenterra, senterrapaint, paintevidence, evidencesubstances, substancesart, artidentify, identifydrugs
Analysis of Polymers and Plastics
2021|Bruker|Brochures and specifications
Analysis of Polymers and Plastics Quality Control & Failure Analysis Innovation with Integrity F T-IR Reliable quality control is essential to achieve a cost-saving production of high quality plastic products. Bruker’s ALPHA II provides a FT-IR spectroscopy-based solution for powerful…
Key words
lumos, lumosquality, qualityalpha, alphaplastics, plasticsinvenio, inveniofailure, failureproduct, productanalysis, analysischemical, chemicalspectroscopy, spectroscopypolymer, polymeroutgassing, outgassingpolymers, polymersplastic, plasticcontrol
Fast and easy analysis of polymeric samples
2021|Bruker|Applications
Application Note AN M84 Fast and easy analysis of polymeric samples fication of an unknown polymer with the aid of the Bruker Introduction polymer library is equally easy; the library search takes just Modern life would be unthinkable without a…
Key words