Measurement of TOC in Electroplating Solution Using TOC-V WS
Applications | | ShimadzuInstrumentation
Organic additives are essential in copper and nickel electroplating solutions to achieve uniform deposit thickness and high surface quality in printed circuit board production. Precise monitoring of these low‐level additives prevents defects and supports consistent manufacturing output. Total organic carbon analysis offers a direct and sensitive approach to quantify organic species in plating baths.
This study evaluates the use of a wet oxidation TOC analyzer for direct measurement of total organic carbon in copper plating solutions. Key goals include establishing calibration procedures for total carbon and inorganic carbon fractions, assessing the need for sample pretreatment, and validating method accuracy through spike recovery experiments.
The procedure involved:
The analysis was conducted using a Shimadzu Wet Oxidation TOC Analyzer TOC‐VWS configured for sequential measurement of TC and IC.
Two commercially relevant copper plating solutions (A and B) were analyzed. Calibration curves demonstrated good linearity through the origin for both TC and IC. Measured values were:
Direct TOC measurement without dilution simplifies sample handling and reduces error potentials. The method reliably detects low‐level organic additives, enabling tighter process control and rapid quality assurance in PCB plating operations.
Advances may include integration of inline TOC monitoring for real‐time process feedback, expansion to other metal plating systems, coupling with chemometric data analysis for predictive maintenance, and development of compact analyzers for on‐site quality checks.
Wet oxidation TOC analysis using the Shimadzu TOC‐VWS provides an effective, accurate, and straightforward approach to quantify organic additives in copper plating baths. High recovery rates and elimination of dilution steps support its adoption for routine quality control in electroplating processes.
TOC
IndustriesEnergy & Chemicals
ManufacturerShimadzu
Summary
Importance of the Topic
Organic additives are essential in copper and nickel electroplating solutions to achieve uniform deposit thickness and high surface quality in printed circuit board production. Precise monitoring of these low‐level additives prevents defects and supports consistent manufacturing output. Total organic carbon analysis offers a direct and sensitive approach to quantify organic species in plating baths.
Study Objectives and Overview
This study evaluates the use of a wet oxidation TOC analyzer for direct measurement of total organic carbon in copper plating solutions. Key goals include establishing calibration procedures for total carbon and inorganic carbon fractions, assessing the need for sample pretreatment, and validating method accuracy through spike recovery experiments.
Methodology
The procedure involved:
- Preparation of calibration standards for total carbon (TC) using potassium hydrogen phthalate at 0 and 2 mgC/L
- Preparation of inorganic carbon (IC) standards using sodium carbonate and sodium hydrogen carbonate at 0 and 1 mgC/L
- Measurement of undiluted copper plating solutions containing approximately 13% copper sulfate without additional pretreatment
- Calculation of TOC by subtracting IC from TC
Instrumentation
The analysis was conducted using a Shimadzu Wet Oxidation TOC Analyzer TOC‐VWS configured for sequential measurement of TC and IC.
Key Results and Discussion
Two commercially relevant copper plating solutions (A and B) were analyzed. Calibration curves demonstrated good linearity through the origin for both TC and IC. Measured values were:
- Solution A: TC 0.443 mgC/L, IC 0.152 mgC/L, resulting TOC 0.291 mgC/L
- Solution B: TC 0.375 mgC/L, IC 0.123 mgC/L, resulting TOC 0.252 mgC/L
Benefits and Practical Applications
Direct TOC measurement without dilution simplifies sample handling and reduces error potentials. The method reliably detects low‐level organic additives, enabling tighter process control and rapid quality assurance in PCB plating operations.
Future Trends and Potential Applications
Advances may include integration of inline TOC monitoring for real‐time process feedback, expansion to other metal plating systems, coupling with chemometric data analysis for predictive maintenance, and development of compact analyzers for on‐site quality checks.
Conclusion
Wet oxidation TOC analysis using the Shimadzu TOC‐VWS provides an effective, accurate, and straightforward approach to quantify organic additives in copper plating baths. High recovery rates and elimination of dilution steps support its adoption for routine quality control in electroplating processes.
Reference
- Shimadzu Application News No O31 Measurement of TOC in Electroplating Solution Using TOC‐VWS
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Measurement of TOC in Electroplating Solution by TOC-L CSH
2013|Shimadzu|Applications
LAAN-A-TC-E021 Application News O45 Total Organic Carbon Analysis Measurement of TOC in Electroplating Solution by TOC-LCSH No. Plating solution that is used in the nickel and copper electroplating process typically contains organic additives. Because these organic additives can affect plating…
Key words
toc, tocplating, platingnickel, nickelsolution, solutionelectroplating, electroplatingmgc, mgcmeasurement, measurementphthalate, phthalateanalyzer, analyzerused, useddilution, dilutionoxidation, oxidationfactor, factorvalue, valueunused
Measurement of TOC in Perfluorochemical (PFOS) Solution
2015|Shimadzu|Applications
LAAN-A-TC-E031 Application News O54 Total Organic Carbon Analysis Measurement of TOC in Perfluorochemical (PFOS) Solution No. Perfluorooctane sulfonate (PFOS) is a man-made organic straight-chain compound consisting of an 8-carbon chain with fluorine atoms bonded to each carbon, and a sulfonic…
Key words
toc, tocmgc, mgcmeasurement, measurementpfos, pfosanalyzer, analyzervalue, valuecombustion, combustioncarbonate, carbonateoxidation, oxidationcatalytic, catalyticwet, wethydrogen, hydrogensolution, solutionorganic, organicperfluorochemical
Recovery Test for Sodium Dodecylbenzenesulfonate by TOC-V WS
|Shimadzu|Applications
SHIMADZU APPLICATION NEWS ● TOTAL ORGANIC CARBON ANALYSIS No. LAAN-A-TC-E011 O34 Recovery Test for Sodium Dodecylbenzenesulfonate by TOC-VWS The test method for organic carbon is listed in the General Tests and Assays section of the Japanese Pharmacopeia (JP). For this…
Key words
dodecylbenzenesulfonate, dodecylbenzenesulfonatetoc, tocvws, vwssodium, sodiumcarbon, carbonmgc, mgcorganic, organicjapanese, japanesesolution, solutioncapable, capablemeasurement, measurementrecovery, recoverysatisfying, satisfyingmeasuring, measuringtest
Recovery Test for Sodium Dodecylbenzenesulfonate by TOC-L CPH
2012|Shimadzu|Applications
LAAN-A-TC-E018 Application News O42 Total Organic Carbon Analysis Recovery Test for Sodium Dodecylbenzenesulfonate by TOC-LCPH No. In the test method for total organic carbon as listed in the General Tests, Processes and Apparatus section of the Japanese Pharmacopeia (JP), it…
Key words
dodecylbenzenesulfonate, dodecylbenzenesulfonatetoc, tocsodium, sodiummgc, mgcmeasurement, measurementaqueous, aqueouscarbon, carbonphthalate, phthalatedistilled, distilledvalue, valuesolution, solutionanalyzer, analyzerhydrogen, hydrogenlcph, lcphcatalytic