Direct swab method procedure using TOC-V and SSM-5000A

Technical notes |  | ShimadzuInstrumentation
TOC
Industries
Manufacturer
Shimadzu

Summary

Significance of Direct Swab TOC Analysis in Cleaning Validation


Effective detection of residual carbon on equipment surfaces is essential in pharmaceutical batch processing to prevent cross-contamination and ensure product purity during cleaning validation processes.

Objectives and Scope of the Application Note


This document outlines a rapid surface sampling approach using glass fiber swabs and direct total organic carbon (TOC) measurement with the Shimadzu TOC-V series coupled to the SSM-5000A module.

Methodology and Instrumentation


The technique employs carbon-free quartz microfibre swabs to collect residues from production surfaces directly. After heat-treatment to remove background carbon, swabs are wetted or loaded with calibration standards and inserted into the SSM-5000A sampling boat. The TOC-V analyzer with solid sampling module allows seamless switching between aqueous and solid measurements via a valve and uses dual NDIR tandem cells—a long path for high sensitivity in aqueous samples and a short path optimized for solids.

Key Results and Discussion


Validation experiments demonstrated detection limits down to 1 µg/L total carbon for swabs processed with the high-sensitivity long cell. Recovery of glucose standards confirmed linear response across the tested range. The direct analysis avoided solvent extraction steps, reducing sample preparation time and potential contamination.

Benefits and Practical Applications


  • Eliminates solvent extraction for swab samples.
  • Enhances sensitivity for low-level residues.
  • Streamlines cleaning validation workflows.
  • Facilitates rapid decision-making in quality control.

Future Trends and Opportunities


Integration with automated cleaning stations, miniaturization of sampling modules, and advanced data analytics for predictive maintenance represent potential developments. Expanding the method to diverse industries, including food and biotechnology, can further broaden applicability.

Conclusion


The direct swab TOC procedure using TOC-V and SSM-5000A offers a robust, sensitive, and efficient means to verify cleaning efficacy on production surfaces, supporting regulatory compliance and product safety.

Reference


Shimadzu Application Note SCA 130 030

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Swab test of E. Coli cells
Swab test of E. Coli cells
2006|Shimadzu|Others
NEWS 03.2006 GB 20.09.2006 10:38 Uhr Seite 15 APPLICATION Shimadzu News 3/2006 Swab test of E. Coli cells ous samples (analysis time is approximately 4 minutes). Figure 1: TOC-V CSH with ASI-V and SSM-5000A Additional solid sample module enables swab…
Key words
swab, swabtoc, tocswabs, swabsboat, boatcleaning, cleaningultrapure, ultrapureanalysis, analysiscell, cellsample, samplecan, canpetrie, petrievalidation, validationceramic, ceramiccose, coseblank
Cleaning Validation by TOC Analyzer with Easy Wiper
LAAN-A-TC-E035 Application News No. Total Organic Carbon Analysis Cleaning Validation by TOC Analyzer with Easy Wiper O60 In the pharmaceutical industry, "cleaning validation" where residues on manufacturing equipment are quantitated to verify whether the amount is within tolerance, has been…
Key words
swab, swabwiper, wipercombustion, combustionvaseline, vaselinesampling, samplinginsoluble, insolublerinderon, rinderontoc, tocointment, ointmentswabbed, swabbedtranexamic, tranexamicisopropylantipyrine, isopropylantipyrinemethod, methodwater, watercleaning
Swab / Direct Combustion Carbon Analysis of Drug Residue by TOC-VCPH
LAAN-A-O-E014 A pplication News Total Organic Carbon Analysis N o . O37 Swab / Direct Combustion Carbon Analysis of Drug Residue by TOC-VCPH Cleaning validation for production equipment at pharmaceutical facilities is an evaluation examination that is conducted to prevent…
Key words
swab, swabvcph, vcphcarbon, carbontoc, tocmeasurement, measurementcombustion, combustionµgc, µgcvalue, valuedrug, drugsubstance, substancedirect, directconducted, conductedresult, resultdistilled, distillednews
Cleaning Validation by TOC Analyzer
LAAN-A-TC-E017 Total Organic Carbon Analysis No.O41 Cleaning Validation by TOC Analyzer To ensure quality control and safety in manufacturing facilities within the pharmaceutical industry, it is important that cleaning validation be conducted following the cleaning of production-related equipment. Cleaning validation…
Key words
ointment, ointmentgentacin, gentacinswab, swabtranexamic, tranexamicisopropylantipyrine, isopropylantipyrinetoc, tocrinderon, rinderonmeasurement, measurementsampling, samplinginsoluble, insolublenifedipine, nifedipineanhydrous, anhydrouswater, watersubstances, substancesanalyzer