QUANTIFICATION OF COMPLEX POLYCYCLIC AROMATIC HYDROCARBONS OR PETROLEUM OILS IN WATER WITH CARY ECLIPSE FLUORESCENCE SPECTROPHOTOMETER ACCORDING TO ASTM D 5412-93 (2000)

Applications | 2013 | Agilent TechnologiesInstrumentation
Fluorescence spectroscopy
Industries
Environmental
Manufacturer
Agilent Technologies

Summary

Importance of the topic


Quantifying polycyclic aromatic hydrocarbons (PAHs) and petroleum oils in water is critical for environmental monitoring, regulatory compliance and human health protection. These contaminants pose carcinogenic risks and require sensitive, selective analytical methods. The ASTM D5412 standard defines a fluorescence‐based approach for accurate measurement of complex PAH mixtures in aqueous samples.

Objectives and overview of the study


This study evaluated the performance of an Agilent Cary Eclipse Fluorescence Spectrophotometer to implement ASTM D5412-93 (2000). Key goals were:
  • Test calibration, precision, limits of detection and quantification for individual PAH analytes (naphthalene, benzo[a]pyrene, anthracene).
  • Compare pipetting approaches and optimize sample preparation.
  • Assess method recovery for individual standards and mixed PAH solutions.

Used Instrumentation


  • Agilent Cary Eclipse Fluorescence Spectrophotometer with quartz cuvettes (1 cm path length).
  • Analytical balance (Ohaus Explorer Pro).
  • Volumetric flasks (A-class borosilicate glass).
  • Automatic pipettes (Eppendorf with Biopur tips), Hamilton microliter syringes.
  • Millipore Milli-Q water system and HPLC-grade cyclohexane.
  • Reagents: naphthalene, anthracene, benzo[a]pyrene standards, nitric acid.

Methodology


Sample and calibration solutions were prepared in cyclohexane according to ASTM D5412. Solvent and water blanks were evaluated to confirm <5% blank intensity. Calibration ranges:
  • Naphthalene: 0.2–100 µg/mL at Ex 250 nm, Em 324 nm.
  • Benzo[a]pyrene: 0.005–0.25 µg/mL at Ex 268 nm, Em 405 nm.

For each level, three replicate spectra were recorded. Slit widths were optimized to improve sensitivity. Limits of detection (LOD) and quantification (LOQ) were calculated using 3σ and 10σ criteria. Precision was divided into repeatability and day-to-day components by variance analysis.

Main Results and Discussion


Key performance data:
  • Naphthalene (Ex 250 nm): LOD 0.13 µg/mL, LOQ 0.5 µg/mL; R²=0.9985.
  • Naphthalene (Ex 268 nm): LOD 0.5 µg/mL, R²=0.9902.
  • Benzo[a]pyrene (narrow slits): LOD 0.06 µg/mL, R²=0.9944.
  • Benzo[a]pyrene (wider slits): LOD 0.005 µg/mL, R²=0.9984.
  • Relative standard deviations were below 0.2 for low‐level replicates and improved at higher concentrations.
  • Recoveries for individual PAHs ranged from 96% to 104%, within ASTM acceptance criteria.
  • Mixed-PAH recoveries were within ±20% across 0.03–0.5 µg/mL total concentration.

Benefits and Practical Applications


The Agilent Cary Eclipse system meets ASTM D5412 requirements, offering:
  • High sensitivity enabling sub-microgram detection of priority PAHs.
  • Robust precision and accuracy for routine water quality analysis.
  • Rapid, cost-effective workflow using standard lab accessories.

Future Trends and Potential Uses


Emerging developments may include:
  • Integration with automated sample extraction and flow-through cells.
  • Enhanced multiplexing for simultaneous detection of multiple PAHs.
  • Application of chemometric methods to deconvolute complex mixtures.
  • Miniaturized and portable fluorimetric systems for field monitoring.

Conclusion


The Cary Eclipse Fluorescence Spectrophotometer successfully implements ASTM D5412-93 for the quantification of PAHs in water. It delivers excellent sensitivity, precision, and recovery for both individual analytes and complex mixtures, providing a reliable solution for environmental and industrial laboratories.

References


  • ASTM D5412-93 (2000) Standard Test Method for Quantification of Complex Polycyclic Aromatic Hydrocarbon Mixtures or Petroleum Oils in Water.
  • ISO/IEC 17025:2005 General requirements for the competence of testing and calibration laboratories.
  • IUPAC (1998) Recommendations for the presentation of results of chemical analysis.
  • Doerffel K. Statistik in der analytischen Chemie. Leipzig: Deutscher Verlag; 1994.
  • Bayrakceken F. Radiative electronic energy transfer studies of naphthalene–biacetyl. Spectrochim. Acta A. 2005;61(6):1069–74.
  • Donaldson DJ et al. Uptake and reaction of atmospheric organic vapours on organic films. Faraday Discuss. 2005;130:227–39.

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
AGILENT WATER ANALYSIS SOLUTIONS - Application Notebook
Application Notebook AGILENT WATER ANALYSIS SOLUTIONS SAFEGUARD PRECIOUS RESOURCES Confidence means safeguarding our precious water, air, and soil with the widest range of high-quality, high-throughput environmental analysis solutions. Agilent Technologies combines more than 40 years of industry collaboration with the…
Key words
notebook, notebookbenzo, benzopass, passfld, fldagilent, agilentfluoranthene, fluoranthenetechnologies, technologiespositive, positivecounts, countsresponses, responsesaldicarb, aldicarbfenamiphos, fenamiphosacquisition, acquisitionsulfoxide, sulfoxidedup
Spectrofluorophotometer RF-6000
Spectrofluorophotometer RF-6000
2025|Shimadzu|Brochures and specifications
C125-E009E Spectrofluorophotometer RF-6000 St ri v i n g f or t he Ultimate in Ac curac y and E a s y Oper ab ility The RF-6000 spectrofluorophotometer has been newly developed by combining Shimadzu’s long-cultivated expertise and cutting-edge…
Key words
fluorescence, fluorescencespectrofluorophotometer, spectrofluorophotometerquantum, quantumspectrum, spectrumcell, cellmeasurement, measurementexcitation, excitationintensity, intensitysphere, sphereefficiency, efficiencycorrection, correctionfluorescent, fluorescentintegrating, integratingyield, yieldholder
AIR/EMISSION - Product Catalogue ALS EUROPE
AIR/EMISSION - Product Catalogue ALS EUROPE
||Brochures and specifications
AIR/EMISSION Product Catalogue ALS EUROPE Right Solutions • Right Partner www.alsglobal.eu Content Air �������������������������������������������������������������������������������������������������������������������������������������������������������������������������������������� 7 Overview of pollutants determined in ALS ����������������������������������������������������������������������������������������������������������������������������������������������7 Ambient air ������������������������������������������������������������������������������������������������������������������������������������������������������������������������ 8 Sulphur dioxide in absorption solution ����������������������������������������������������������������������������������������������������������������������������������������������������8 Sulphur dioxide from Radiello* sampling tube ���������������������������������������������������������������������������������������������������������������������������������������8 Nitrogen dioxide from Radiello*…
Key words
parameter, parameterfluorantene, fluoranteneloq, loqpcb, pcbsampling, samplingcdf, cdfbenzo, benzocatalogue, catalogueright, rightcdd, cddteq, teqair, airortho, ortholor, lorvoc
The Linear Dynamic Range and Limits of Detection of Fluorescein using the Agilent Cary Eclipse Fluorescence Spectrophotometer
The Linear Dynamic Range and Limits of Detection of Fluorescein using the Agilent Cary Eclipse Fluorescence Spectrophotometer Technical Overview Introduction Fluorescein is one of the most widely used fluorescent labels and is very well characterized. Fluorescein has absorption and emission…
Key words
fluorescence, fluorescencefluorescein, fluoresceincary, caryemission, emissionexcitation, excitationfluorescein, fluoresceineclipse, eclipsespectrophotometer, spectrophotometerfluorescence, fluorescencefilter, filterlinear, lineardynamic, dynamiclog, logslit, slitlimit
Other projects
GCMS
LCMS
Follow us
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike