Hollow Cathode Lamps – Yesterday, Today and Tomorrow
Technical notes | 2010 | Agilent TechnologiesInstrumentation
Atomic absorption spectroscopy relies on hollow cathode lamps to provide sharp, intense emission lines essential for sensitive and selective elemental analysis. Improvements in lamp design, such as boosted discharge and multi-element lamps, enhance analytical performance in various applications including environmental monitoring, industrial quality control, and research.
This article reviews the fundamental principles of hollow cathode lamps, traces their historical development, and highlights advances in lamp designs. Focus is placed on multi-element hollow cathode lamps and boosted discharge lamps that increase signal intensity and reduce self absorption.
The principles of lamp operation are described
Construction features that ensure sharp spectral output are summarized. The history of lamp development from early CSIRO work to modern production at Agilent is outlined.
Boosted discharge lamps demonstrate significant enhancements in line intensity, for example over fivefold for the Au 242.8 nm line and notable gains for Pb, Ag, Cd, and Zn lines. Multi-element lamps allow simultaneous determination of up to six elements with acceptable sensitivity. Calibration curves show reduced curvature at high absorbances, indicating lower self absorption. FAQs address lamp lifetime, spectral interferences, and handling.
Enhanced lamp designs improve detection limits and reduce photon shot noise, benefiting flame and furnace AA methods. Multi-element lamps and fast sequential instruments enable efficient, cost effective analysis for metals in diverse matrices. Longer lamp life and stable output reduce maintenance.
Ongoing development aims to expand multi-element combinations, improve boosted lamp efficiency for additional lines, and integrate novel lamp materials. Enhanced automation and higher throughput AA systems will drive broader industrial and research adoption.
Hollow cathode lamps remain indispensable for atomic absorption spectroscopy. Advances in boosted discharge and multi-element lamps deliver greater sensitivity and versatility, supporting rapid, reliable elemental analysis across applications.
AAS
IndustriesManufacturerAgilent Technologies
Summary
Importance of the Topic
Atomic absorption spectroscopy relies on hollow cathode lamps to provide sharp, intense emission lines essential for sensitive and selective elemental analysis. Improvements in lamp design, such as boosted discharge and multi-element lamps, enhance analytical performance in various applications including environmental monitoring, industrial quality control, and research.
Objectives and Study Overview
This article reviews the fundamental principles of hollow cathode lamps, traces their historical development, and highlights advances in lamp designs. Focus is placed on multi-element hollow cathode lamps and boosted discharge lamps that increase signal intensity and reduce self absorption.
Methodology
The principles of lamp operation are described
- Sputtering of cathode material by ionized fill gas
- Excitation of sputtered atoms and emission of spectral lines
Construction features that ensure sharp spectral output are summarized. The history of lamp development from early CSIRO work to modern production at Agilent is outlined.
Used Instrumentation
- Conventional hollow cathode lamps with neon or argon fill gas
- Boosted discharge hollow cathode lamps with independent excitation circuit
- SpectrAA 220FS fast sequential atomic absorption spectrometer
Main Results and Discussion
Boosted discharge lamps demonstrate significant enhancements in line intensity, for example over fivefold for the Au 242.8 nm line and notable gains for Pb, Ag, Cd, and Zn lines. Multi-element lamps allow simultaneous determination of up to six elements with acceptable sensitivity. Calibration curves show reduced curvature at high absorbances, indicating lower self absorption. FAQs address lamp lifetime, spectral interferences, and handling.
Benefits and Practical Applications
Enhanced lamp designs improve detection limits and reduce photon shot noise, benefiting flame and furnace AA methods. Multi-element lamps and fast sequential instruments enable efficient, cost effective analysis for metals in diverse matrices. Longer lamp life and stable output reduce maintenance.
Future Trends and Opportunities
Ongoing development aims to expand multi-element combinations, improve boosted lamp efficiency for additional lines, and integrate novel lamp materials. Enhanced automation and higher throughput AA systems will drive broader industrial and research adoption.
Conclusion
Hollow cathode lamps remain indispensable for atomic absorption spectroscopy. Advances in boosted discharge and multi-element lamps deliver greater sensitivity and versatility, supporting rapid, reliable elemental analysis across applications.
References
- J V Sullivan and A Walsh Spectrochimica Acta 26B 721 1965
- J V Sullivan Progress in Analytical Atomic Spectroscopy 4 311 1981
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Practical Multi-Element Hollow Cathode Lamps for Atomic Absorption Spectrometry
2010|Agilent Technologies|Applications
Practical Multi-Element Hollow Cathode Lamps for Atomic Absorption Spectrometry Application Note Atomic Absorption Authors Introduction Douglas Shrader Multi-element hollow cathode lamps have traditionally been considered inferior to single-element lamps for spectrochemical analysis by atomic absorption (AA). This publication examines some…
Key words
lamps, lampselement, elementlamp, lampmulti, multicathode, cathodesingle, singlehollow, hollowmisconception, misconceptionboosted, boostedmetal, metalagree, agreelines, linesemission, emissionsteel, steelhave
Hollow Cathode Lamps
2018|Agilent Technologies|Technical notes
White Paper Hollow Cathode Lamps Competitive Comparison The hollow cathode lamp is a discharge lamp designed for use as a spectral line source with atomic absorption (AA) spectrometers. A single or multi-element hollow cathode lamp is required for each element…
Key words
lamp, lamplamps, lampssgm, sgmcathode, cathodegrinm, grinmhollow, hollowvarsal, varsalheraeus, heraeusphotron, photronagilent, agilentabsorbance, absorbancelifetime, lifetimehrs, hrsoperating, operatingintensity
Features and Operation of Hollow Cathode Lamps and Deuterium Lamps
2010|Agilent Technologies|Technical notes
Features and Operation of Hollow Cathode Lamps and Deuterium Lamps Application Note Atomic Absorption Authors Abstract K. Brodie and S. Neate This paper discusses the parameters which influence the operation of hollow cathode and deuterium lamps. Examples are given of…
Key words
lamp, lampcathode, cathodehollow, hollowdeuterium, deuteriumlamps, lampsintensity, intensitynoise, noisespectral, spectralemission, emissionenvelope, envelopeabsorbance, absorbancegetter, getterwarm, warmabsorption, absorptionline
Evaluation of High Intensity Lamps for AAS
2010|Agilent Technologies|Applications
Evaluation of High Intensity Lamps for AAS Application Note Atomic Absorption Authors Introduction Jonathan H. Moffett The trend over the three decades since atomic absorption spectrometry (AAS) was introduced as a very successful analytical technique [1] has been to measure…
Key words
ultraa, ultraalamp, lamplamps, lampstmf, tmfcrm, crmtmdw, tmdwhollow, hollowgraphs, graphsemission, emissionsputtering, sputteringw’length, w’lengthphotomultiplier, photomultipliercharacteristic, characteristiccathode, cathodeselenium