Methods for Environmental Samples
Applications | 2010 | Agilent TechnologiesInstrumentation
Trace element monitoring in environmental matrices is critical for assessing pollution, enforcing regulatory limits and protecting public health. Graphite furnace atomic absorption spectrometry provides the sensitivity and selectivity required to quantify elements at parts-per-billion levels in samples such as water, soil and sediment. Optimized furnace programs, advanced background correction and matrix modification techniques help laboratories meet stringent detection limits and ensure data accuracy under diverse sample conditions.
This application study was designed to develop and validate graphite furnace atomic absorption methods for nine priority trace elements identified by the US EPA: arsenic, selenium, cadmium, lead, thallium, antimony, silver, beryllium and chromium. Key goals included achieving contract required detection limits, demonstrating long-term method stability, minimizing chemical and spectral interferences and maximizing sample throughput for routine environmental monitoring.
A suite of Varian SpectrAA-300/400 Zeeman spectrometers equipped with graphite furnace modules and programmable sample dispensers was employed. Hot Inject technology enabled rapid drying stages. Photron super lamps served as element sources where applicable. Method features included:
The optimized methods routinely met or exceeded EPA Superfund contract required detection limits for all elements. Calibration linearity was excellent across the specified ranges. Long-term precision and accuracy were demonstrated by repeated analysis of blanks, check standards, certified soil digests and environmental water samples. Key findings included:
The validated GFAAS protocols enable environmental and contract laboratories to:
Advances in furnace design, automated sampling and software-guided method optimization will further streamline trace element workflows. Emerging matrix modifier chemistries may extend the range of analytes and sample types. Coupling GFAAS with hyphenated techniques or miniaturized furnaces could enable field-deployable solutions for on-site environmental screening.
This study demonstrates that graphite furnace atomic absorption spectrometry, when combined with platform atomization, Zeeman background correction and carefully selected matrix modifiers, achieves precise, accurate and rapid quantification of key environmental trace elements. Laboratories can reliably meet regulatory requirements while maintaining high throughput and data quality.
AAS
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Significance of Topic
Trace element monitoring in environmental matrices is critical for assessing pollution, enforcing regulatory limits and protecting public health. Graphite furnace atomic absorption spectrometry provides the sensitivity and selectivity required to quantify elements at parts-per-billion levels in samples such as water, soil and sediment. Optimized furnace programs, advanced background correction and matrix modification techniques help laboratories meet stringent detection limits and ensure data accuracy under diverse sample conditions.
Objectives and Study Overview
This application study was designed to develop and validate graphite furnace atomic absorption methods for nine priority trace elements identified by the US EPA: arsenic, selenium, cadmium, lead, thallium, antimony, silver, beryllium and chromium. Key goals included achieving contract required detection limits, demonstrating long-term method stability, minimizing chemical and spectral interferences and maximizing sample throughput for routine environmental monitoring.
Methodology and Used Instrumentation
A suite of Varian SpectrAA-300/400 Zeeman spectrometers equipped with graphite furnace modules and programmable sample dispensers was employed. Hot Inject technology enabled rapid drying stages. Photron super lamps served as element sources where applicable. Method features included:
- Platform atomization to separate sample from furnace wall and reduce matrix chemical effects
- Zeeman background correction to eliminate non-specific absorption and correct spectral interferences
- Element-specific matrix modifiers approved by US EPA SW-846 to stabilize volatile analytes and suppress matrix effects
- Peak area absorbance measurement for enhanced precision
- Direct calibration over practical ranges without recourse to standard additions
Main Results and Discussion
The optimized methods routinely met or exceeded EPA Superfund contract required detection limits for all elements. Calibration linearity was excellent across the specified ranges. Long-term precision and accuracy were demonstrated by repeated analysis of blanks, check standards, certified soil digests and environmental water samples. Key findings included:
- Arsenic and selenium required nickel matrix modifiers and benefited from premixed low-level standards to reduce curvature at microquantities
- Thallium performance improved markedly by combining palladium modifier with citric acid as reducing agent
- Cadmium, lead and antimony methods attained recoveries between 98 and 105 percent, with relative standard deviations below 5 percent on real samples
- Beryllium and chromium did not require modifiers, relying on platform or wall atomization, respectively
- Typical single-sample analysis times were under two minutes, supporting high throughput
Benefits and Practical Applications
The validated GFAAS protocols enable environmental and contract laboratories to:
- Comply with federal regulatory detection limits and QA/QC requirements
- Process large sample loads rapidly with minimal manual intervention
- Deliver accurate results across diverse matrices without time-consuming standard additions
- Maintain method robustness against common matrix interferences encountered in water and soil analyses
Future Trends and Potential Applications
Advances in furnace design, automated sampling and software-guided method optimization will further streamline trace element workflows. Emerging matrix modifier chemistries may extend the range of analytes and sample types. Coupling GFAAS with hyphenated techniques or miniaturized furnaces could enable field-deployable solutions for on-site environmental screening.
Conclusion
This study demonstrates that graphite furnace atomic absorption spectrometry, when combined with platform atomization, Zeeman background correction and carefully selected matrix modifiers, achieves precise, accurate and rapid quantification of key environmental trace elements. Laboratories can reliably meet regulatory requirements while maintaining high throughput and data quality.
References
- US EPA SW-846 Test Methods for Evaluating Solid Waste, approved matrix modifier guidelines
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
The Determination of the Priority Pollutant Metals Using the CRA-90 Carbon Rod Atomizer
2010|Agilent Technologies|Applications
The Determination of the Priority Pollutant Metals Using the CRA-90 Carbon Rod Atomizer Application Note Atomic Absorption Authors Introduction Lucinda M. Voth Over the past few years, increasing concern has been shown over the potentially detrimental effects of trace metal…
Key words
ash, ashcra, craramp, rampeffluent, effluentacetic, aceticcalibration, calibrationdry, drydrinking, drinkingwastewatr, wastewatrera, eraadditions, additionswater, wateracid, acidcorrected, correctedaqueous
Arsenic in Natural Waters by Graphite Furnace Atomic Absorption using EPA Method 200.9.
2016|Thermo Fisher Scientific|Applications
Application Note: 40851 Arsenic in Natural Waters by Graphite Furnace Atomic Absorption using EPA Method 200.9. Introduction Key Words • Arsenic • Atomic Absorption • EPA 200.9 • GFAAS • Water • Environment • Zeeman Key Benefits • Advanced spectrometer…
Key words
furnace, furnacegraphite, graphitearsenic, arsenicsolaar, solaarpass, passwater, wateratomic, atomicmethod, methodabsorption, absorptionnatural, naturalatomize, atomizesignal, signalused, usedqcs, qcsstandard
Determination of Tin by Hydride Generation AA
2010|Agilent Technologies|Applications
Determination of Tin by Hydride Generation AA Application Note Atomic Absorption Author Introduction Lucinda M Beach Hydride generation techniques are used routinely for the determination of As, Se, and Sb in atomic absorption and atomic emission spectroscopy. Low level concentrations…
Key words
hydride, hydridegeneration, generationtin, tingraphite, graphitefurnace, furnacenormal, normalcysteine, cysteinelockheed, lockheedvga, vgachannel, channelreductant, reductantacid, acidmethodology, methodologyyes, yesdetermination
Agilent Inorganic Certified Reference Materials and Standards
2021|Agilent Technologies|Brochures and specifications
Agilent Inorganic Certified Reference Materials and Standards Your essential resource for single, multi-element and metallo-organic standards for spectroscopy and IC instrumentation Table of Contents Introduction3 Multi-Element Agency Method Standards 32 Why you need Agilent Standards 3 EPA Method 200.7 standards…
Key words
thallium, thalliumstandards, standardscobalt, cobaltbarium, bariumcadmium, cadmiumberyllium, berylliumelement, elementselenium, seleniumchromium, chromiummanganese, manganesenickel, nickelarsenic, arsenicicp, icpvanadium, vanadiumzinc