High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System – Discrete Sampling (ISIS–DS)
Applications | 2009 | Agilent TechnologiesInstrumentation
High-speed and reliable multi-element analysis of environmental samples is critical for regulatory compliance and resource monitoring. The integration of advanced sampling accessories with ICP-MS instrumentation addresses the growing demand for fast, accurate, and low-volume analyses.
This study evaluates the performance of the Agilent 7500cx ICP-MS equipped with the newly configured Integrated Sample Introduction System–Discrete Sampling accessory in helium collision mode. The aim was to achieve USEPA-compliant analysis of over 30 elements in approximately 75 seconds per sample, with sub-ppb detection limits and effective interference removal.
A discrete sampling approach using a 300 µL sample loop and a high-speed peristaltic uptake pump was implemented with the Cetac ASX-520 autosampler. The tubing between the injection valve and the nebulizer was minimized to reduce dispersion, yielding a 15-second stabilization delay. Spectrum mode acquisition with three replicates was performed under helium collision cell conditions, avoiding the need for gas switching and enabling rapid run times.
The optimized method achieved a sample-to-sample time of 75 seconds and consumed 2.2 mL of sample per analysis. A sequence of 216 samples was completed in 4 hours and 29 minutes. Internal standard recoveries remained within ±10% across all injections, and calibration check recoveries stayed within ±5%, complying with USEPA Method 200.8 criteria. Analysis of NIST 1643e certified reference water showed recoveries within ±10% of certified values and relative standard deviations near 1% for most elements. Washout testing demonstrated greater than 99.9% signal reduction in the first rinse blank for all analytes, including high-matrix elements such as Na, Mg, Al, K, and Fe.
The ISIS-DS configuration offers ultra-high throughput with minimal sample volume, making it ideal for laboratories facing high analysis demands or limited sample availability. Integrated software control, reduced carryover risk, and low waste generation contribute to enhanced operational efficiency and data reliability in environmental, QA/QC, and industrial settings.
Future developments may include integration of multiple reaction gases for expanded interference control, use of hydrogen mode for ultra-trace elements such as Se, further automation of sample preparation workflows, and miniaturization of sampling accessories for on-site field measurements.
The reconfigured ISIS-DS accessory on the Agilent 7500cx ICP-MS enables rapid, accurate, and interference-free environmental analyses that fully comply with USEPA standards while significantly reducing sample and reagent consumption.
ICP/MS
IndustriesEnvironmental
ManufacturerAgilent Technologies
Summary
Significance of the Topic
High-speed and reliable multi-element analysis of environmental samples is critical for regulatory compliance and resource monitoring. The integration of advanced sampling accessories with ICP-MS instrumentation addresses the growing demand for fast, accurate, and low-volume analyses.
Objectives and Study Overview
This study evaluates the performance of the Agilent 7500cx ICP-MS equipped with the newly configured Integrated Sample Introduction System–Discrete Sampling accessory in helium collision mode. The aim was to achieve USEPA-compliant analysis of over 30 elements in approximately 75 seconds per sample, with sub-ppb detection limits and effective interference removal.
Methodology and Instrumentation
A discrete sampling approach using a 300 µL sample loop and a high-speed peristaltic uptake pump was implemented with the Cetac ASX-520 autosampler. The tubing between the injection valve and the nebulizer was minimized to reduce dispersion, yielding a 15-second stabilization delay. Spectrum mode acquisition with three replicates was performed under helium collision cell conditions, avoiding the need for gas switching and enabling rapid run times.
Used Instrumentation
- Agilent 7500cx ICP-MS with Octopole Reaction System in helium collision mode
- Integrated Sample Introduction System–Discrete Sampling (ISIS-DS) accessory
- Cetac ASX-520 autosampler with 0.8 mm internal diameter probe
- Standard glass concentric nebulizer and spray chamber
Main Results and Discussion
The optimized method achieved a sample-to-sample time of 75 seconds and consumed 2.2 mL of sample per analysis. A sequence of 216 samples was completed in 4 hours and 29 minutes. Internal standard recoveries remained within ±10% across all injections, and calibration check recoveries stayed within ±5%, complying with USEPA Method 200.8 criteria. Analysis of NIST 1643e certified reference water showed recoveries within ±10% of certified values and relative standard deviations near 1% for most elements. Washout testing demonstrated greater than 99.9% signal reduction in the first rinse blank for all analytes, including high-matrix elements such as Na, Mg, Al, K, and Fe.
Benefits and Practical Applications
The ISIS-DS configuration offers ultra-high throughput with minimal sample volume, making it ideal for laboratories facing high analysis demands or limited sample availability. Integrated software control, reduced carryover risk, and low waste generation contribute to enhanced operational efficiency and data reliability in environmental, QA/QC, and industrial settings.
Future Trends and Applications
Future developments may include integration of multiple reaction gases for expanded interference control, use of hydrogen mode for ultra-trace elements such as Se, further automation of sample preparation workflows, and miniaturization of sampling accessories for on-site field measurements.
Conclusion
The reconfigured ISIS-DS accessory on the Agilent 7500cx ICP-MS enables rapid, accurate, and interference-free environmental analyses that fully comply with USEPA standards while significantly reducing sample and reagent consumption.
References
- Agilent Technologies, High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System–Discrete Sampling ISIS–DS, Application Note 5990-3678EN, March 2009.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Maximizing productivity for high matrix sample analysis using the Agilent 7900 ICP-MS with ISIS 3 discrete sampling system
2014|Agilent Technologies|Applications
Maximizing productivity for high matrix sample analysis using the Agilent 7900 ICP-MS with ISIS 3 discrete sampling system EPA 6020A compliant analysis in less than 90 seconds per sample Application note Authors Environmental Kazuo Yamanaka1 and Steve Wilbur2 1 Agilent…
Key words
ppb, ppbmatrix, matrixmean, meanccv, ccvrecovery, recoveryllicv, llicvspike, spikellccv, llccvccb, ccbisis, isisrpd, rpdbenefits, benefitsistd, istdmode, modeisotope
Maximizing Productivity in High Matrix Samples using the Agilent 7700x ICP-MS with ISIS Discrete Sampling
2010|Agilent Technologies|Technical notes
Maximizing Productivity in High Matrix Samples using the Agilent 7700x ICP-MS with ISIS Discrete Sampling EPA 6020A Compliant Analysis in Less Than 2 Minutes per Sample Application Note Environmental Authors Abstract Steve Wilbur Discrete sampling, also known as flow injection,…
Key words
isis, isismean, meanspike, spikeppb, ppbmatrix, matrixrecovery, recoverydiscrete, discreteaerosol, aerosolelement, elementuptake, uptakereducing, reducingmuch, muchduplicates, duplicatesrsd, rsdhigh
Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS
2007|Agilent Technologies|Technical notes
Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS Application Environmental Author Steven Wilbur Agilent Technologies, Inc. 3380 146th Place, SE, Suite 300 Bellevue, WA 98007 USA Abstract The new Agilent 7500cx with Octopole Reaction System (ORS) is…
Key words
polyatomic, polyatomicics, icsmode, modehelium, heliuminterferences, interferencesacq, acqstabilization, stabilizationusepa, usepappt, pptelements, elementsccv, ccvgas, gassuffer, sufferuptake, uptakeccb
Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3
2017|Agilent Technologies|Applications
Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3 Application Author Environmental Tetsuo Kubota Agilent Technologies, Japan Introduction Regulations concerning activities such as land use, waste discharge consents, and remediation of former industrial sites require routine…
Key words
ppb, ppbistd, istdllicv, llicvsoil, soilmatrix, matrixsample, sampleloop, loopspike, spikeplasma, plasmaicv, icvelement, elementccv, ccvpreset, presetrecovery, recoveryrsd