Photometric accuracy of internal DRAs in the UV-Vis region using certified potassium dichromate solution

Brochures and specifications | 2011 | Agilent TechnologiesInstrumentation
UV–VIS spectrophotometry
Industries
Manufacturer
Agilent Technologies

Summary

Significance of the Topic


The reliability of absorbance measurements in UV-Vis spectrophotometry underpins quantitative analyses across pharmaceuticals, environmental testing, and materials research. Verifying photometric accuracy, especially when using integrated accessories like diffuse reflectance attachments, ensures consistency among instruments and compliance with regulatory standards.

Objectives and Study Overview


This study evaluates the photometric accuracy of Agilent’s New Generation Cary 4000, 5000, and 6000i spectrophotometers equipped with the internal Diffuse Reflectance Accessory (DRA). By measuring a certified potassium dichromate standard, the work aims to demonstrate instrument performance against traceable reference values.

Methodology and Instrumentation


Instrumentation:
  • Cary 4000/5000/6000i spectrophotometer with internal DRA
  • DRA cuvette holder (part number 0210187900)
  • Certified potassium dichromate solution (Starna kit 9910085200)

Protocol:
  • Warm up instrument for 1 hour
  • Auto-calibrate DRA via the Validate application
  • Align and mount the cuvette holder
  • Set scan parameters: 200–800 nm, 60 nm/min, 1 nm interval, 0.1 s signal averaging, 1.5 nm SBW, zero/baseline on
  • Perform zero/baseline correction with blank
  • Measure absorbance of standard solution
  • Compare results to certificate values

Main Results and Discussion


Absorbance at key wavelengths matched certified values within ±0.004 absorbance units. For example, at 350 nm the certified absorbance of 0.638 was measured as 0.642 (difference 0.004). All deviations remained below the 0.01 tolerance, confirming excellent system accuracy. Data demonstrate that the integrated DRA does not compromise photometric performance.

Benefits and Practical Applications


Reliable photometric accuracy with integrated accessories streamlines UV-Vis workflows by:
  • Enabling direct comparison of results across multiple instruments
  • Meeting pharmacopoeial and regulatory validation requirements
  • Reducing need for external calibration fixtures
  • Supporting quality control in pharmaceutical, environmental, and industrial labs

Future Trends and Possibilities


Advances may include automated self-validation routines, integration of smart diagnostics with AI-driven trend analysis, and broader wavelength range accessories. Enhanced accessory interoperability and cloud-based performance tracking will further simplify compliance and maintenance.

Conclusion


This validation confirms that Agilent’s Cary 4000/5000/6000i spectrophotometers with internal DRA deliver high photometric accuracy in the UV-Vis region. The system meets stringent tolerance criteria using a certified potassium dichromate standard, supporting diverse analytical applications.

References


Solution Validation Test kit, part number 9910085200

Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.

Downloadable PDF for viewing
 

Similar PDF

Toggle
Linear dynamic range of the Cary 4000, 5000, 6000i: internal diffuse reflectance accessories
Linear dynamic range of the Cary 4000, 5000, 6000i: internal diffuse reflectance accessories Data Sheet Introduction/Theory The photometric accuracy and linearity of a spectrophotometer defines its ability to measure an absorbance that can be directly related to a compound of…
Key words
dra, dradynamic, dynamicpermanganate, permanganatereflectance, reflectancephotometric, photometricrange, rangelinear, linearwide, wideabsorbance, absorbanceinternal, internalrang, rangturbid, turbidlinearity, linearityorientation, orientationcary
Agilent Cary 4000/5000/6000i Series UV-VIS-NIR Spectrophotometers
Agilent Cary 4000/5000/6000i Series UV-VIS-NIR Spectrophotometers
2022|Agilent Technologies|Brochures and specifications
Unrivaled. Precise. Consistent. Agilent Cary 4000/5000/6000i Series UV-VIS-NIR Spectrophotometers Unrivaled Agilent is your premier resource and partner for molecular spectroscopy. The world-renowned Cary product line, encompasses FTIR, UV-Vis-NIR and Fluorescence, offering you a comprehensive range of molecular spectroscopy solutions. Answers…
Key words
accessory, accessoryreflectance, reflectancewavelength, wavelengthdiffuse, diffusenir, nirabs, absholder, holdercary, caryangle, anglemeasurements, measurementsingaas, ingaascoatings, coatingssample, sampleoptical, opticalfilm
Linear dynamic range of the Cary 4000, 5000, 6000i: external diffuse reflectance accessories
Linear dynamic range of the Cary 4000, 5000, 6000i: external diffuse reflectance accessories Data Sheet Introduction/Theory The photometric accuracy and linearity of a spectrophotometer defines its ability to measure an absorbance that can be directly related to a compound of…
Key words
dynamic, dynamicdra, drapermanganate, permanganatephotometric, photometricrange, rangelinear, linearwide, wideabsorbance, absorbancerang, rangturbid, turbidlinearity, linearityorientation, orientationcary, caryexcellent, excellentreflectance
UV-VISIBLE HARDWARE OPERATIONAL QUALIFICATION
UV-VISIBLE HARDWARE OPERATIONAL QUALIFICATION
2012|Agilent Technologies|Others
UV-VISIBLE HARDWARE OPERATIONAL QUALIFICATION Agilent Enterprise Edition Compliance Services Standard OQ Test Specifications for UV-Visible Spectrophotometers The following tables describe the core hardware testing specifications for UV-Visible systems. See the corresponding attachment for a description of the optional tests available…
Key words
spectrophotometers, spectrophotometersrationale, rationalevisible, visibletest, testenterprise, enterpriseedition, editionaccuracy, accuracycontinued, continueddesign, designsetpoints, setpointsservices, serviceslimits, limitscompliance, compliancewavelength, wavelengthmodels
Other projects
GCMS
LCMS
Follow us
FacebookLinkedInYouTube
More information
WebinarsAbout usContact usTerms of use
LabRulez s.r.o. All rights reserved. Content available under a CC BY-SA 4.0 Attribution-ShareAlike