Moisture analysis in fertilizer products
Applications | 2020 | MetrohmInstrumentation
Moisture content in fertilizers critically influences product quality, shelf life, and compliance with legal limits. Accurate, rapid moisture measurement safeguards production efficiency and prevents microbial growth or caking during storage.
This study demonstrates the application of near-infrared spectroscopy (NIRS) for fast, reagent-free moisture analysis in various fertilizer products. It compares NIRS results with coulometric Karl Fischer titration to validate accuracy.
A set of fertilizer samples with moisture levels ranging from 0.12 % to 3.82 % was analyzed. The NIRS DS2500 Solid Analyzer collected reflectance spectra using a rotating large sample cup to minimize inhomogeneity. Data acquisition and chemometric modeling were performed with Vision Air 2.0 software. Reference moisture values were established by coulometric KF titration coupled with an oven. Pre-processing employed a second derivative filter, and a partial least squares (PLS) algorithm was calibrated and validated via cross-validation.
The NIRS model achieved a coefficient of determination R² = 0.98 when predicting moisture versus KF titration. The standard error of calibration (SEC) was 0.29 % and the standard error of cross-validation (SECV) was 0.32 %, indicating strong agreement and model robustness across different fertilizer types.
Advances may include inline process integration for real-time moisture monitoring, expanded calibration libraries for additional fertilizer parameters, miniature portable spectrometers for field use, and machine-learning algorithms to further improve prediction accuracy.
NIRS with the DS2500 Solid Analyzer provides a rapid, reliable, and environmentally friendly approach to moisture analysis in fertilizers, matching the accuracy of conventional KF titration and streamlining quality control workflows.
No formal references were provided in the source document.
NIR Spectroscopy
IndustriesEnergy & Chemicals
ManufacturerMetrohm
Summary
Importance of the Topic
Moisture content in fertilizers critically influences product quality, shelf life, and compliance with legal limits. Accurate, rapid moisture measurement safeguards production efficiency and prevents microbial growth or caking during storage.
Objectives and Study Overview
This study demonstrates the application of near-infrared spectroscopy (NIRS) for fast, reagent-free moisture analysis in various fertilizer products. It compares NIRS results with coulometric Karl Fischer titration to validate accuracy.
Methodology and Instrumentation
A set of fertilizer samples with moisture levels ranging from 0.12 % to 3.82 % was analyzed. The NIRS DS2500 Solid Analyzer collected reflectance spectra using a rotating large sample cup to minimize inhomogeneity. Data acquisition and chemometric modeling were performed with Vision Air 2.0 software. Reference moisture values were established by coulometric KF titration coupled with an oven. Pre-processing employed a second derivative filter, and a partial least squares (PLS) algorithm was calibrated and validated via cross-validation.
Used Instrumentation
- DS2500 Solid Analyzer (400–2500 nm range; robust design for dust, moisture, vibration, temperature fluctuations)
- Vision Air 2.0 Complete spectroscopy software (data management, method development, routine analysis)
- Large sample cup for homogeneous measurement of powders and granulates
Main Results and Discussion
The NIRS model achieved a coefficient of determination R² = 0.98 when predicting moisture versus KF titration. The standard error of calibration (SEC) was 0.29 % and the standard error of cross-validation (SECV) was 0.32 %, indicating strong agreement and model robustness across different fertilizer types.
Benefits and Practical Applications
- Analysis time under one minute per sample
- No chemical reagents or waste generation
- High reproducibility and ease of use in laboratory and production settings
- Automated multi-sample capability enhances throughput
Future Trends and Opportunities
Advances may include inline process integration for real-time moisture monitoring, expanded calibration libraries for additional fertilizer parameters, miniature portable spectrometers for field use, and machine-learning algorithms to further improve prediction accuracy.
Conclusion
NIRS with the DS2500 Solid Analyzer provides a rapid, reliable, and environmentally friendly approach to moisture analysis in fertilizers, matching the accuracy of conventional KF titration and streamlining quality control workflows.
Reference
No formal references were provided in the source document.
Content was automatically generated from an orignal PDF document using AI and may contain inaccuracies.
Similar PDF
Moisture analysis in caprolactam
2021|Metrohm|Applications
AN-NIR-077 Moisture analysis in caprolactam Moisture control without chemical waste Summary Caprolactam is an important polymer used for the production of Nylon 6, which is the base material for industrial fibers. Other application areas for caprolactam are in the resin…
Key words
metrohm, metrohmsimple, simplemoisture, moistureday, daygranulates, granulatesmanagement, managementvalidation, validationsecv, secvcaprolactam, caprolactamsql, sqlionenstrasse, ionenstrassepls, plscup, cuppowders, powdersinterfaces
Quality Control of Polyamides
2020|Metrohm|Applications
AN-NIR-060 Quality Control of Polyamides Determination of viscosity, functional groups, and moisture within one minute using NIR Spectroscopy Summary Functional group and viscosity analysis (ASTM D789) of polyamides can be a lengthy and challenging process due to the sample’s limited…
Key words
metrohm, metrohmerror, errormerit, meritfigures, figurestitration, titrationcross, crossstandard, standardresult, resultvalidation, validationgroup, groupend, endcarboxyl, carboxylsimple, simplevalue, valuenir
Quality control of polyvinyl alcohol
2021|Metrohm|Applications
AN-NIR-076 Quality control of polyvinyl alcohol Improved polymer quality control with NIR spectroscopy Summary Polyvinyl alcohol (PVA) is used in a variety of medical products (e.g. eye drops) due to its low toxicity, low protein adhesion, and film forming properties.…
Key words
metrohm, metrohmsimple, simplealcoholysis, alcoholysisday, daygranulates, granulatesmanagement, managementvalidation, validationsecv, secvsql, sqlpolyvinyl, polyvinylionenstrasse, ionenstrassepls, plscup, cuppowders, powdersinterfaces
Quality control of Ammonium Nitrate
2020|Metrohm|Applications
AN-NIR-064 Quality control of Ammonium Nitrate Rapid and non-destructive moisture determination Summary Specialty chemicals have to fulfill multiple quality requirements. One of these quality parameters, which can be found in almost all certificates of analysis and specifications, is the moisture…
Key words
metrohm, metrohmper, persimple, simpleday, daynir, nircosts, costsgranulates, granulateserror, errormanagement, managementsql, sqlequipment, equipmentsoftware, softwaremerit, meritionenstrasse, ionenstrassevision